Calibration of small grating spectrometers from 166 to 600 cm-1
Open Access
- 1 May 1962
- journal article
- Published by National Institute of Standards and Technology (NIST) in Journal of Research of the National Bureau of Standards Section A: Physics and Chemistry
- Vol. 66A (3) , 223-225
- https://doi.org/10.6028/jres.066a.021
Abstract
In order to provide standards for the calibration of small grating spectrometers over the region from 166 to 600 cm−1, tracings of the spectrum of atmospheric water vapor are presented. The lines are identified and tabulated. Wavenumbers obtained from energy levels derived from the best available high-resolution spectra are given, together with an indication of their relative reliability. The best lines are believed accurate to ±0.03 cm−1.Keywords