Strain distribution in the proximal femur with flexible composite and metallic femoral components under axial and torsional loads
- 1 May 1993
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 27 (5) , 575-585
- https://doi.org/10.1002/jbm.820270504
Abstract
This study investigated strain distribution changes in the proximal femur after implantation of a flexible composite femoral component (carbon composite material, modulus of elasticity = 18.6 GPa), a titanium alloy implant (E = 100 GPa), and a stainless steel implant (E = 200 GPa). Transverse as well as longitudinal strain was measured using bipolar strain gauges at eight locations on the proximal femur under both physiologic axial (1000 N and 2000 N) and physiologic torsional (10 N‐m and 20 N‐m) loads. Under axial load, longitudinal compressive strain at the calcar region was significantly greater in intact femurs and the carbon composite stem specimens than in the two metal stem specimens. The difference between intact femurs and the carbon composite stem specimens was not significant. Stress shielding in the proximal lateral region of the femur, however, was still apparent even in the carbon composite stem specimens. Without seating of the stem collar on the femoral neck, longitudinal compressive strain was not generated at the calcar region, and transverse tensile strain at this region was increased. With conventional implant design, the stem collar was still necessary even in the flexible composite stem to provide near normal longitudinal compressive strain in the calcar region. Under torsional load, proximal strain in intact femurs was small and the proximal strain levels observed after either carbon composite or titanium alloy stem implantation were greater than strain levels before implantation. It seemed unlikely that torsional stress relief played a significant role in proximal bone loss after total hip arthroplasty. Both longitudinal and transverse strains at the calcar region under torsional load were significantly greater in the carbon composite stem specimens than in both intact femurs and the titanium alloy stem specimens, suggesting that these abnormally high proximal stresses may cause high proximal micromotion of the implant, and even bone fracture. Proximal implant design seems to be of paramount importance with flexible composite femoral components to avoid excessive proximal stress concentration under torsional load and‐to provide near normal strain distribution in the proximal femur. © 1993 John Wiley & Sons, Inc.Keywords
This publication has 22 references indexed in Scilit:
- In vitro analysis of proximal femoral strains using PCA femoral implants and a hip-abductor muscle simulatorThe Journal of Arthroplasty, 1989
- Initial effect of collarless stem stiffness on femoral bone strainThe Journal of Arthroplasty, 1989
- Effect of femoral component section modulus on the stress distribution in the proximal human femurMedical & Biological Engineering & Computing, 1988
- Strains and micromotions of press-fit femoral stem prosthesesJournal of Biomechanics, 1987
- Effects of stem design and material properties on stresses in hip endoprosthesesJournal of Biomedical Engineering, 1987
- Instability of total hip prostheses at rotational stress: A roentgen stereophotogrammetric studyActa Orthopaedica, 1984
- A Measurement of Proximal Femur Strain with Total Hip ArthroplastyJournal of Biomechanical Engineering, 1980
- The influence of design parameters on calcar stresses following femoral head arthroplastyJournal of Biomedical Materials Research, 1980
- Fracture of Femoral Prostheses in Total Hip ReplacementClinical Orthopaedics and Related Research, 1975
- The elastic modulus for boneJournal of Biomechanics, 1974