Quality control in haematology: Report of interlaboratory trials in britain
- 1 November 1969
- Vol. 4 (5678) , 253-256
- https://doi.org/10.1136/bmj.4.5678.253
Abstract
An interlaboratory quality control scheme has been established in Britain by the British Committee for Standards in Haematology. In the first instance this has been confined to haemoglobin, red blood cell count, and packed cell volume. The materials which have been circulated include whole blood, stabilized red cell preparations, lysates, cyanmethaemoglobin solutions, and cyanmethaemoglobin reference preparations. The first two trials have been completed, and there are implications for instrument calibration, dilution techniques, and the use of standards. Periodic interlaboratory trials, at a national level, in conjunction with regular individual intralaboratory quality control procedures, are necessary in order to achieve acceptable levels of accuracy and precision.Keywords
This publication has 4 references indexed in Scilit:
- PRESERVATION OF BLOOD FOR TRANSFUSION .6. EFFECT OF ADDITION OF ADENINE AND INOSINE ON ATP AND POSTTRANSFUSION SURVIVAL OF RED CELLS OF STORED BLOOD1968
- Further Studies on the Effect of Adenine in Blood Preservation*Transfusion, 1967
- A stable standard suspension for red-cell counts.1966
- The Number Plus Method of Quality Control of Laboratory AccuracyAmerican Journal of Clinical Pathology, 1963