Abstract
This document, from the International Society for Clinical Electrophysiology of Vision (ISCEV), presents an updated and revised ISCEV Standard for clinical electroretinography (ERG). The parameters for flash stimulation and background adaptation have been tightened, and responses renamed to indicate the flash strength (in cd·s·m−2). The ISCEV Standard specifies five responses: (1) Dark-adapted 0.01 ERG (rod response); (2) Dark-adapted 3.0 ERG (combined rod–cone response); (3) Dark-adapted 3.0 oscillatory potentials; (4) Light-adapted 3.0 ERG (cone response); (5) Light-adapted 3.0 flicker (30 Hz flicker). An additional Dark-adapted 10.0 ERG or Dark-adapted 30.0 ERG response is recommended.Keywords
This publication has 7 references indexed in Scilit:
- ISCEV guidelines for clinical multifocal electroretinography (2007 edition)Documenta Ophthalmologica, 2007
- ISCEV standard for clinical pattern electroretinography—2007 updateDocumenta Ophthalmologica, 2007
- ISCEV Standard for Clinical Electro-oculography (EOG) 2006Documenta Ophthalmologica, 2006
- Standard for clinical electroretinography (2004 update)Documenta Ophthalmologica, 2004
- Visual evoked potentials standard (2004)Documenta Ophthalmologica, 2004
- Guidelines for calibration of stimulus and recording parameters used in clinical electrophysiology of visionDocumenta Ophthalmologica, 2003