Quality-assurance reports: CDRH RF phantom for hyperthermia systems evaluations
- 1 January 1988
- journal article
- research article
- Published by Taylor & Francis in International Journal of Hyperthermia
- Vol. 4 (1) , 17-23
- https://doi.org/10.3109/02656738809032046
Abstract
The National Cancer Institute (NCI) sponsored clinical evaluations of investigational ''regional'' hyperthermia systems at four clinical institutions. To support this project, the Center for Devices and Radiological Health (CDRH) developed a series of test instruments to evaluate the magnitude and repeatibility of the induced heating by radiofrequency (RF) systems. Data from these institutions using the same model hyperthermia system have been analyzed. After heating, the average temperature from measurements taken at several points in the test phantom at each institution agree within .+-. 0.002.degree. C. These differences are about equal to the measurement uncertainty. Thus, this technique can be used for preclinical evaluation and quality control of the total system operation. After one of the institutions relocated its hyperthermia system, a subsequent set of data showed inconsistencies compared to their earlier data. Investigation traced this to cable loss and power meter interference. From the analysis of the data from the three institutions, the utility of the CDRH RF phantom for hyperthermia systems evaluation is demonstrated.This publication has 3 references indexed in Scilit:
- Development of a Family of RF Helical Coil Applicators Which Produce Transversely Uniform Axially Distributed Heating in Cylindrical Fat-Muscle PhantomsIEEE Transactions on Biomedical Engineering, 1984
- Regional Hyperthermia with an Annular Phased ArrayIEEE Transactions on Biomedical Engineering, 1984
- The UHF and microwave dielectric properties of normal and tumour tissues: variation in dielectric properties with tissue water contentPhysics in Medicine & Biology, 1980