Comparison of the Internal Energy Deposition of Venturi-Assisted Electrospray Ionization and a Venturi-Assisted Array of Micromachined UltraSonic Electrosprays (AMUSE)
- 30 September 2008
- journal article
- research article
- Published by American Chemical Society (ACS) in Journal of the American Society for Mass Spectrometry
- Vol. 19 (9) , 1320-1329
- https://doi.org/10.1016/j.jasms.2008.06.012
Abstract
The internal energy deposition of a Venturi-assisted array of micromachined ultrasonic electrosprays (AMUSE), with and without the application of a DC charging potential, is compared with equivalent experiments for Venturi-assisted electrospray ionization (ESI) using the “survival yield” method on a series of para-substituted benzylpyridinium salts. Under conditions previously shown to provide maximum ion yields for standard compounds, the observed mean internal energies were nearly identical (1.93–2.01 eV). Operation of AMUSE without nitrogen flow to sustain the air amplifier focusing effect generated energetically colder ions with mean internal energies that were up to 39% lower than those for ESI. A balance between improved ion transfer, adequate desolvation, and favorable ion energetics was achieved by selection of optimum operational ranges for the parameters that most strongly influence the ion population: the air amplifier gas flow rate and API capillary temperature. Examination of the energy landscapes obtained for combinations of these parameters showed that a low internal energy region (≤1.0 eV) was present at nitrogen flow rates between 2 and 4 L min−1 and capillary temperatures up to 250°C using ESI (9% of all parameter combinations tested). Using AMUSE, this region was present at nitrogen flow rates up to 2.5 L min−1 and all capillary temperatures (13% of combinations tested). The signal-to-noise (S/N) ratio of the intact p-methylbenzylpyridinium ion obtained from a 5 µM mixture of thermometer compounds using AMUSE at the extremes of the studied temperature range was at least fivefold higher than that of ESI, demonstrating the potential of AMUSE ionization as a soft method for the characterization of labile species by mass spectrometry.Keywords
This publication has 22 references indexed in Scilit:
- Identification of dimethyldioctadecylammonium ion (m/z 550.6) and related species (m/z 522.6, 494.6) as a source of contamination in mass spectrometryJournal of the American Society for Mass Spectrometry, 2008
- Probing the mechanisms of an air amplifier using a LTQ-FT-ICR-MS and fluorescence spectroscopyJournal of the American Society for Mass Spectrometry, 2007
- Quantitative comparison of a flared and a standard heated metal capillary inlet with a voltage-assisted air amplifier on an electrospray ionization linear ion trap mass spectrometerRapid Communications in Mass Spectrometry, 2007
- Analytical Performance of a Venturi-Assisted Array of Micromachined Ultrasonic Electrosprays Coupled to Ion Trap Mass Spectrometry for the Analysis of Peptides and ProteinsAnalytical Chemistry, 2007
- Generation of Highly Charged Peptide and Protein Ions by Atmospheric Pressure Matrix-Assisted Infrared Laser Desorption/Ionization Ion Trap Mass SpectrometryAnalytical Chemistry, 2007
- Gentle Protein Ionization Assisted by High-Velocity Gas FlowAnalytical Chemistry, 2005
- Direct Observation of Sugar-Protein, Sugar-Sugar, and Sugar-Water Complexes by Cold-Spray Ionization Time-of-Flight Mass SpectrometryAngewandte Chemie International Edition in English, 2005
- Analytical Performance of a Venturi Device Integrated into an Electrospray Ionization Fourier Transform Ion Cyclotron Resonance Mass Spectrometer for Analysis of Nucleic AcidsAnalytical Chemistry, 2004
- Incorporation of a Venturi Device in Electrospray IonizationAnalytical Chemistry, 2003
- Laser desorption ionization of proteins with molecular masses exceeding 10,000 daltonsAnalytical Chemistry, 1988