Detection of Metals in the Environment Using a Portable Laser-Induced Breakdown Spectroscopy Instrument
- 1 February 1996
- journal article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 50 (2) , 222-233
- https://doi.org/10.1366/0003702963906519
Abstract
A portable instrument, based on laser-induced breakdown spectroscopy (LIBS), has been developed for the detection of metal contaminants on surfaces. The instrument has a weight of 14.6 kg, fits completely into a small suitcase (46 × 33 × 24 cm), and operates from 115 V ac. The instrument consists of a sampling probe connected to the main analysis unit by electrical and optical cabling. The hand-held probe contains a small laser to generate laser sparks on a surface and a fiber-optic cable to collect the spark light. The collected light is spectrally resolved and detected with the use of a compact spectrograph/CCD detector system. The instrument has been evaluated for the analysis of metals in the environment: Ba, Be, Pb, and Sr in soils; Pb in paint; and Be and Pb particles collected on filters. Detection limits in ppm for metals in soils were 265 (Ba), 9.3 (Be), 298 (Pb), and 42 (Sr). The detection limit for Pb in paint was 0.8% (8000 ppm), corresponding to 0.052 mg/cm2. The higher limit obtained for Pb in paint is attributed to the use of the 220.35-nm Pb(II) line instead of the stronger 405.78-nm Pb(I) line used for soils. Spectral interferences prevented use of the 405.78-nm line to determine Pb in paint. The surface detection limit for Be particles on filters was dependent on particle size and ranged from 21 to 63 ng/cm2. The detection limit for Pb particles on filters was 5.6 μg/cm2.Keywords
This publication has 10 references indexed in Scilit:
- Remote Elemental Analysis by Laser-Induced Breakdown Spectroscopy Using a Fiber-Optic CableApplied Spectroscopy, 1995
- Laser ablation ICP-mass spectrometry determination of Th{sup 230} in soils at the Gunnison, Colorado UMTRA sitePublished by Office of Scientific and Technical Information (OSTI) ,1994
- A High Resolution Portable XRF HgI2 Spectrometer for Field Screening of Hazardous Metal WastesPublished by Springer Nature ,1992
- Reheating of a Laser-Produced Plasma by a Second Pulse LaserApplied Spectroscopy, 1991
- XRF Technique as a Method of Choice for On-Site Analysis of Soil Contaminants and Waste MaterialPublished by Springer Nature ,1990
- Detection of cadmium, lead and zinc in aerosols by laser-induced breakdown spectrometryJournal of Analytical Atomic Spectrometry, 1988
- Direct Detection of Beryllium on Filters Using the Laser SparkApplied Spectroscopy, 1985
- Spectrochemical Analysis of Liquids Using the Laser SparkApplied Spectroscopy, 1984
- A review and tutorial discussion of noise and signal-to-noise ratios in analytical spectrometry—I. Fundamental principles of signal-to-noise ratiosSpectrochimica Acta Part B: Atomic Spectroscopy, 1978
- Methods of Soil AnalysisPublished by Wiley ,1965