Comparison of the Jackson-Rees Circuit, the Pediatric Circle, and the MERA F Breathing System for Pediatric Anesthesia
- 1 September 1996
- journal article
- Published by Wolters Kluwer Health in Anesthesia & Analgesia
- Vol. 83 (3) , 488-492
- https://doi.org/10.1213/00000539-199609000-00008
Abstract
The purpose of this study was to compare the Jackson-Rees circuit with the pediatric circle and MERA F breathing system (MERA F system) for pediatric anesthesia from the viewpoint of work of breathing (WOB). Twenty-three children (2-10 yr old) were studied during spontaneous breathing under endotracheal anesthesia with 4 L/min nitrous oxide, 2 L/min oxygen, and 1% end-tidal concentration of sevoflurane. WOB, inspiratory and expiratory airway resistance, dynamic compliance (CDYN), pressure time product (PTP), and arterial blood gasses were measured in the three circuits. The inspiratory WOB was estimated directly by measuring the esophageal pressure-volume loop using the Campbell technique. In a laboratory study, we measured the compliances of the Jackson-Rees circuit, the pediatric circle, the MERA F system, and the adult circuit. WOB differed among the three circuits (MERA F system > pediatric circle > Jackson-Rees circuit). Inspiratory and expiratory resistances, and arterial carbon dioxide tension in the Jackson-Rees circuit were significantly lower than those of both the pediatric circle and MERA F system. The CDYN and PTP in the MERA F system were significantly higher than those in both the Jackson-Rees circuit and the pediatric circle. The MERA F system had significantly higher compliance than the Jackson-Rees circuit and pediatric circle. It is concluded that the Jackson-Rees circuit is most efficient, the pediatric circle is intermediate, and the MERA F system is the least efficient from the viewpoint of WOB during spontaneous breathing for pediatric anesthesia. (Anesth Analg 1996;83:488-92)Keywords
This publication has 10 references indexed in Scilit:
- Comparison of circle absorber and jacksonrees systems for paediatric anaesthesiaCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1988
- Ventilatory responses to inspiratory mechanical loads in spontaneously breathing children during halothane anaesthesiaActa Anaesthesiologica Scandinavica, 1986
- VENTILATORY RESPONSES TO REBREATHING AND CARBON DIOXIDE INHALATION DURING ANAESTHESIA IN CHILDRENBritish Journal of Anaesthesia, 1985
- CHANGE IN THE WORK OF BREATHING IMPOSED BY FIVE ANAESTHETIC BREATHING SYSTEMSBritish Journal of Anaesthesia, 1983
- Lung Mechanics in Sitting and Horizontal Body PositionsChest, 1983
- Rebreathing and coaxial circuits: A comparison of the bain and mera FCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1981
- Stochastic differential equations with reflecting boundary condition in convex regionsHiroshima Mathematical Journal, 1979
- A Comparison of Respiratory Effort of Infants Anesthetized with Several Adult and Pediatric SystemsAnesthesiology, 1966
- Relation between work output of respiratory muscles and end-tidal CO2 tensionJournal of Applied Physiology, 1963
- Anaesthesia in the NewbornBMJ, 1950