The in vitro hemodynamic characteristics of a variety of mechanical and tissue heart valve designs used during the past two decades were investigated in the aortic position under pulsatile flow conditions. The following valve designs were studied: Starr-Edwards ball and cage (model 1260), Björk-Shiley tilting disc (convexo-concave model), Medtronic-Hall tilting disc, St. Jude Medical bileaflet, Carpentier-Edwards porcine and pericardial (models 2625, 2650 and 2900), Hancock porcine (models 250 and 410) and Ionescu-Shiley standard pericardial. The Starr-Edward ball and cage, Björk-Shiley tilting disc, Carpentier-Edwards porcine (model 2625) and Ionescu-Shiley standard pericardial valves were designed prior to 1975, while the Medtronic-Hall tilting disc, St. Jude Medical bileaflet, Hancock porcine (model 250), Hancock II porcine (model 410), Carpentier-Edwards porcine (model 2650) and Carpentier-Edwards pericardial (model 2900) valves were designed after 1975. The pressure drop results indicated that the valves designed prior to 1975 had performance indices of 0.30 to 0.45, whereas the valves designed after 1975 had performance indices of 0.40 to 0.70. The regurgitant volumes were higher for the mechanical designs (5.0 to 11.0 cm3/beat) compared to the tissue bioprostheses (1.0 to 5.0 cm3/ beat). Two-dimensional laser Doppler anemometry studies indicated that the valves designed after 1975 tended to create more centralized flow fields, with reduced levels of turbulent shear stresses. However, none of the current valve designs is ideal: they all create areas of stasis and/or regions of low velocity reverse flow; and regions of elevated turbulent shear stresses that are capable of causing sub-lethal and/or lethal damage to the formed elements of blood.Keywords
This publication has 10 references indexed in Scilit:
- Evaluation of Prosthetic Heart Valves by Doppler Flow ImagingEchocardiography, 1986
- In vitro hemodynamic characteristics of tissue bioprostheses in the aortic positionThe Journal of Thoracic and Cardiovascular Surgery, 1986
- In vitro pulsatile flow velocity and shear stress measurements in the vicinity of mechanical mitral heart valve prosthesesJournal of Biomechanics, 1986
- Turbulent shear stress measurements in the vicinity of aortic heart valve prosthesesJournal of Biomechanics, 1986
- Pulsatile flow past aortic valve bioprostheses in a model human aortaJournal of Biomechanics, 1984
- Laser anemometry measurements of pulsatile flow past aortic valve prosthesesJournal of Biomechanics, 1983
- In vitro velocity measurements in the vicinity of aortic prosthesesJournal of Biomechanics, 1979
- The Björk-Shiley aortic prosthesis: flow characteristics, thrombus formation and tissue overgrowth.Circulation, 1978
- Choosing a substitute cardiac valve: Type, size, surgeonThe American Journal of Cardiology, 1976