Fourier Deconvolution of Overlapping Line Pairs in Inductively Coupled Plasma-Atomic Emission Spectrometry
- 1 January 1989
- journal article
- research article
- Published by SAGE Publications in Applied Spectroscopy
- Vol. 43 (1) , 96-103
- https://doi.org/10.1366/0003702894202049
Abstract
A deconvolution procedure utilizing Fourier transformation has been developed to reduce line overlap in ICP-AES. Line broadening is caused by physical processes and by instrumental broadening. Convenient deconvolution, however, turns out to be restricted to broadening common to the emission lines in the spectral window, i.e., to instrumental broadening. Deconvolution for the “true” instrumental broadening function and for a Gaussian approximation to this function yields similar results, but the former allows for fast automated data processing with regard to any spectral region and sample composition. A straightforward procedure is reported for the determination of this function independent of wavelength. At the present noise level, a twofold reduction in linewidth can be achieved for emission lines having a small physical width in comparison to the instrumental width. With data acquired from both a high- and a medium-resolution monochromator, results from overlapping line pairs show linear analytical curves and improved detection limits. Due to the decrease in signal-to-noise ratio on deconvolution, the detection limits measured for isolated lines cannot be attained.Keywords
This publication has 19 references indexed in Scilit:
- Detection limit including selectivity as a criterion for line selection in trace analysis using inductively coupled plasma-atomic emission spectrometry (ICP-AES)—a tutorial treatment of a fundamental problem of AESSpectrochimica Acta Part B: Atomic Spectroscopy, 1987
- Inductively coupled plasma Fourier transform spectrometry: a new analytical technique? potentials and problems. Plenary lectureJournal of Analytical Atomic Spectrometry, 1987
- Inductively coupled plasmas: line widths and shapes, detection limits and spectral interferences. An integrated picture. Plenary lectureJournal of Analytical Atomic Spectrometry, 1987
- Computer simulation of emission from analyte atoms and ions excited in an inductively coupled plasmaSpectrochimica Acta Part B: Atomic Spectroscopy, 1986
- High-resolution spectroscopy using an echelle spectrometer with predisperser-II. Analytical optimization for inductively coupled plasma atomic emission spectrometrySpectrochimica Acta Part B: Atomic Spectroscopy, 1984
- Some Observations on the Resolution Enhancement of Spectral Data by the Method of Self-DeconvolutionApplied Spectroscopy, 1983
- Fourier Self-Deconvolution: A Method for Resolving Intrinsically Overlapped BandsApplied Spectroscopy, 1981
- Corrections for spectral interferences in optical emission spectrometry with special reference to the RF inductively coupled plasmaSpectrochimica Acta Part B: Atomic Spectroscopy, 1976
- A Simple and Versatile Fourier Domain Digital FilterApplied Spectroscopy, 1976
- Digital data handling of spectra utilizing Fourier transformationsAnalytical Chemistry, 1972