Transformation of Whole-Head MEG Recordings Between Different Sensor Positions / Transformation von Ganzkopf-MEG-Messungen zwischen verschiedenen Sensorpositionen
- 1 January 2002
- journal article
- Published by Walter de Gruyter GmbH in Biomedizinische Technik/Biomedical Engineering
- Vol. 47 (3) , 59-62
- https://doi.org/10.1515/bmte.2002.47.3.59
Abstract
In this work, a method is presented for the transformation of MEG recordings to a standard sensor position. For the case of an 148 channel magnetometer array, the algorithm was evaluated using simulations as well as phantom head recordings. It turned out that the method is very robust and yields accurate results regardless of the position of the field generator, even with large differences between original and standard (target) sensor positions and high noise. The method can be applied as a prerequisite for comparing or averaging recordings of different subjects or sessions.Keywords
This publication has 3 references indexed in Scilit:
- Comparability of measurement results obtained with multi-SQUID-systems of different sensor configurationsIEEE Transactions on Applied Superconductivity, 1997
- Transformation of multichannel magnetocardiographic signals to standard grid formIEEE Transactions on Biomedical Engineering, 1995
- Magnetoencephalography: A tool for functional brain imagingBrain Topography, 1992