Data Interpretation for Coupled Molecular Weight Sensitive Detectors in Sec: Interdetector Transport Time
- 1 January 1992
- journal article
- research article
- Published by Taylor & Francis in Journal of Liquid Chromatography
- Vol. 15 (1) , 39-69
- https://doi.org/10.1080/10826079208018808
Abstract
The time required for polymer molecules to pass from one detector to another is a critical parameter for interpretation of multidetector size exclusion chromatography data. A method of utilizing numerical optimization to determine this quantity is presented. The chromatography system included a differential refractive index detector (DRI) in combination with a low-angle laser light scattering photometer (LALLS) and a differential viscometer (DV). The optimization method requires a conventional calibration curve (in terms of intrinsic viscosity rather than molecular weight if the DV detector is used) and the raw data from analysis of a broad molecular weight distribution linear homopolymer standard. A cubic polynomial and a least square cubic spline were used to fit the conventional calibration curve. In interpretation of the DV data, it was necessary to use a least square cubic spline to fit the conventional calibration curve. Results obtained are “effective” in that they intrinsically provide agreement with various molecular weight measures of the standard. Some evidence shows that they are slightly lower than estimates from experimental methods available.Keywords
This publication has 8 references indexed in Scilit:
- Quantitative Size Exclusion Chromatography: Assessing New DevelopmentsJournal of Liquid Chromatography, 1990
- Measurement of Interdetector Volume in Size Exclusion/Low-Angle Laser Light Scattering ExperimentsJournal of Liquid Chromatography, 1990
- Evaluation of a Thermal Pulse Flowmeter for Quantitative Size Exclusion ChromatographyJournal of Liquid Chromatography, 1990
- Size Exclusion Chromatography Calibration Assessment Utilizing Coupled Molecular Weight DetectorsJournal of Liquid Chromatography, 1990
- An Experimental Evaluation of a New Commercial Viscometric Detector for Size-Exclusion Chromatography (SEC) Using Linear and Branched PolymersJournal of Liquid Chromatography, 1987
- Applicability of the Mayo–Lewis equation to high conversion copolymerization of styrene and methylmethacrylateJournal of Polymer Science: Polymer Chemistry Edition, 1984
- Polymer analysis by fractionation with on-line light scattering detectorsChromatographia, 1982
- Measurement of the Dead Volume between Concurrent Detectors in Gel Permeation ChromatographyJournal of Liquid Chromatography, 1982