Equations for Measuring Blood Flow by External Monitoring of Radioisotopes
- 1 April 1965
- journal article
- research article
- Published by Wolters Kluwer Health in Circulation Research
- Vol. 16 (4) , 309-321
- https://doi.org/10.1161/01.res.16.4.309
Abstract
It is possible by external monitoring alone to measure blood flow per unit volume through any vascular bed accessible to external monitoring of the radioactive tracer if the tracer input function or if the quantity of tracer entering the system is known. The methods proposed exploit the fact that the mean transit time through the system is the flow per unit volume of distribution of tracer. The equations are free of assumptions that require solutions as exponentials or any other specified frequency function of transit times. If the input function is known, measurement of the concentration of tracer in blood leaving the part sensed by the external detector, combined with external detection of radioactivity, can lead to measurement of absolute blood flow.Keywords
This publication has 5 references indexed in Scilit:
- Regional Blood Flow of the Cerebral Cortex Determined by Krypton85Acta Physiologica Scandinavica, 1962
- THEORY OF THE USE OF ARTERIOVENOUS CONCENTRATION DIFFERENCES FOR MEASURING METABOLISM IN STEADY AND NON-STEADY STATES*Journal of Clinical Investigation, 1961
- A METHOD FOR THE CONTINUOUS MEASUREMENT OF CEREBRAL BLOOD FLOW IN MAN BY MEANS OF RADIOACTIVE KRYPTON (Kr79)*Journal of Clinical Investigation, 1960
- On the Theory of the Indicator-Dilution Method for Measurement of Blood Flow and VolumeJournal of Applied Physiology, 1954
- Measurement of regional circulation by the local clearance of radioactive sodiumAmerican Heart Journal, 1949