Oligoclonal bands in cerebrospinal fluid detected by PhastSystem isoelectric focusing
- 1 January 1990
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 36 (1) , 123-125
- https://doi.org/10.1093/clinchem/36.1.123
Abstract
Pharmacia's "PhastSystem" for semi-automated isoelectric focusing (IEF) in thin precast polyacrylamide gels (PAGE) was found to be as sensitive as high-resolution protein electrophoresis (HRPE) in agarose gels and conventional PAGE-IEF for detection of oligoclonal banding (OB) in concentrated cerebrospinal fluid (CSF) samples. Both PhastSystem IEF and HRPE revealed OB in CSF from eight of nine multiple sclerosis patients and four of 10 patients with various types of infection of the central nervous system as opposed to only two of 70 patients with miscellaneous neuropsychiatric disorders. The PhastSystem also frequently detected OB in silver-stained, unconcentrated CSF from patients with multiple sclerosis.This publication has 3 references indexed in Scilit:
- Oligoclonal bands in cerebrospinal fluids: significance of corresponding bands in serum for diagnosis of multiple sclerosis.Clinical Chemistry, 1988
- Analysis for cerebrospinal fluid proteins by isoelectric focusing on polyacrylamide gel: methodological aspects and normal values, with special reference to the alkaline region.Clinical Chemistry, 1978
- Biochemical findings in multiple sclerosisZeitschrift für Neurologie, 1977