Dosimetric verification of a commercial inverse treatment planning system
- 1 January 1999
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 44 (2) , 463-478
- https://doi.org/10.1088/0031-9155/44/2/013
Abstract
A commercial three-dimensional (3D) inverse treatment planning system, Corvus (Nomos Corporation, Sewickley, PA), was recently made available. This paper reports our preliminary results and experience with commissioning this system for clinical implementation. This system uses a simulated annealing inverse planning algorithm to calculate intensity-modulated fields. The intensity-modulated fields are divided into beam profiles that can be delivered by means of a sequence of leaf settings by a multileaf collimator (MLC). The treatments are delivered using a computer-controlled MLC. To test the dose calculation algorithm used by the Corvus software, the dose distributions for single rectangularly shaped fields were compared with water phantom scan data. The dose distributions predicted to be delivered by multiple fields were measured using an ion chamber that could be positioned in a rotatable cylindrical water phantom. Integrated charge collected by the ion chamber was used to check the absolute dose of single- and multifield intensity modulated treatments at various spatial points. The measured and predicted doses were found to agree to within 4% at all measurement points. Another set of measurements used a cubic polystyrene phantom with radiographic film to record the radiation dose distribution. The films were calibrated and scanned to yield two-dimensional isodose distributions. Finally, a beam imaging system (BIS) was used to measure the intensity-modulated x-ray beam patterns in the beam's-eye view. The BIS-measured images were then compared with a theoretical calculation based on the MLC leaf sequence files to verify that the treatment would be executed accurately and without machine faults. Excellent correlation (correlation coefficients > or = 0.96) was found for all cases. Treatment plans generated using intensity-modulated beams appear to be suitable for treatment of irregularly shaped tumours adjacent to critical structures. The results indicated that the system has potential for clinical radiation treatment planning and delivery and may in the future reduce treatment complexity.Keywords
This publication has 18 references indexed in Scilit:
- An optimized leaf-setting algorithm for beam intensity modulation using dynamic multileaf collimatorsPhysics in Medicine & Biology, 1998
- Quality assurance for dynamic multileaf collimator modulated fields using a fast beam imaging systemMedical Physics, 1997
- An iterative filtered backprojection inverse treatment planning algorithm for tomotherapyInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Realization and verification of three-dimensional conformal radiotherapy with modulated fieldsInternational Journal of Radiation Oncology*Biology*Physics, 1994
- 3-D planning and delivery system for optimized conformal therapyInternational Journal of Radiation Oncology*Biology*Physics, 1992
- A finite‐size pencil beam model for photon dose calculations in three dimensionsMedical Physics, 1992
- A review of electronic portal imaging devices (EPIDs)Medical Physics, 1992
- Methods of image reconstruction from projections applied to conformation radiotherapyPhysics in Medicine & Biology, 1990
- Optimization of stationary and moving beam radiation therapy techniquesRadiotherapy and Oncology, 1988
- A protocol for the determination of absorbed dose from high‐energy photon and electron beamsMedical Physics, 1983