Pulse inversion Doppler: a new method for detecting nonlinear echoes from microbubble contrast agents
- 1 March 1999
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control
- Vol. 46 (2) , 372-382
- https://doi.org/10.1109/58.753026
Abstract
A novel technique for the selective detection of ultrasound contrast agents, called pulse inversion Doppler, has been developed. In this technique, a conventional Doppler or color Doppler pulse sequence is modified by inverting every second transmit pulse. Either conventional or harmonic Doppler processing is then performed on the received echoes. In the resulting Doppler spectra, Doppler shifts from linear and nonlinear scattering are separated into two distinct regions that can be analyzed separately or combined to estimate the ratio of nonlinear to linear scattering from a region of tissue. The maximum Doppler shift that can be detected is 1/2 the normal Nyquist limit. This has the advantage over conventional harmonic Doppler that it can function over the entire bandwidth of the echo signal, thus achieving superior spatial resolution in the Doppler image. In vitro measurements comparing flowing agent and cellulose particles suggest that pulse inversion Doppler can provide 3 to 10 dB more agent to tissue contrast than harmonic imaging with similar pulses. Similar measurements suggest that broadband pulse inversion Doppler can provide up to 16 dB more contrast than broadband conventional Doppler. Nonlinear propagation effects limit the maximum contrast obtainable with both harmonic and pulse inversion Doppler techniques.Keywords
This publication has 13 references indexed in Scilit:
- Characteristics of contrast agents and 2D imagingPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2002
- Predicting acoustic response of a microbubble population for contrast imagingPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2002
- Pulse inversion Doppler: a new method for detecting nonlinear echoes from microbubble contrast agentsPublished by Institute of Electrical and Electronics Engineers (IEEE) ,2002
- Color flow mappingUltrasound in Medicine & Biology, 1997
- An extended autocorrelation method for estimation of blood velocityIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1997
- Finite amplitude distortion-based inhomogeneous pulse echo ultrasonic imagingIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1997
- Imaging instrumentation for ultrasound contrast agentsPublished by Springer Nature ,1997
- An axial velocity estimator for ultrasound blood flow imaging, based on a full evaluation of the Doppler equation by means of a two-dimensional autocorrelation approachIEEE Transactions on Ultrasonics, Ferroelectrics, and Frequency Control, 1995
- Nonlinear acoustical response of coated microbubbles in diagnostic ultrasoundPublished by Institute of Electrical and Electronics Engineers (IEEE) ,1994
- Description of broad-band pulsed Doppler ultrasound processing using the two-dimensional Fourier transformUltrasonic Imaging, 1991