Closed-Loop Control of an Anesthesia Delivery System: Development and Animal Testing
- 1 June 1987
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Biomedical Engineering
- Vol. BME-34 (6) , 437-443
- https://doi.org/10.1109/TBME.1987.326078
Abstract
An anesthesia system which integrates closed-loop control of ventilation, oxygen, nitrous oxide, and anesthetic agent delivery into a closed breathing circuit is described. Breathing circuit volume is regulated by controlling the sum of oxygen and nitrous oxide flow, and oxygen concentration in the breathing circuit is regulated by controlling the ratio of oxygen flow to the sum of oxygen and nitrous oxide flow. End-tidal anesthetic agent concentration is regulated by controlling the agent delivery, and end-tidal carbon dioxide concentration is regulated by controlling the ventilation. After tuning the ventilation and anesthetic delivery controllers in preliminary trials in ten dogs, the system was tested and evaluated in five dogs. All control loops were stable and exhibited time responses to step changes in setpoint or external disturbances which were clinically acceptable. The system makes available the advantages of closed-system anesthesia without encumbering the anesthesiologist with the control tasks associated with the technique.Keywords
This publication has 5 references indexed in Scilit:
- SERVO-CONTROLLED CLOSED-CIRCUIT ANAESTHESIABritish Journal of Anaesthesia, 1983
- UPTAKE OF ENFLURANE: A STUDY OF THE VARIABILITY BETWEEN PATIENTSBritish Journal of Anaesthesia, 1983
- Parameter-optimized ControllersPublished by Springer Nature ,1981
- Instrumentation for Measuring Continuous Oxygen Consumption of Surgical PatientsIEEE Transactions on Biomedical Engineering, 1977
- Computer control of respiration and anaesthesiaMedical & Biological Engineering & Computing, 1973