Objective Symmetry Detector Method for Gammaencephalography: I. Physical characteristics
- 1 January 1975
- journal article
- Published by Taylor & Francis in Acta Radiologica: Therapy, Physics, Biology
- Vol. 14 (1) , 63-80
- https://doi.org/10.3109/02841867509132263
Abstract
A symmetry detector method for gammanecephalography is described. It is based on two detectors, collimated by square tubes with large apertures. Increase ratios and side difference ratios should be used as evaluation parameters in clinical applications and are therefore analysed theoretically and experimentally from measurements on a phantom system. The effect of energy discrimination is analyzed and found to be small over a wide range of the lower discrimination level setting. The chosen performance of the measuring routine used made errors caused by the electronic devices negligible in relation to the loss of precision occurring from counting statistics.Keywords
This publication has 4 references indexed in Scilit:
- Gamma-Encephalography after Ten Years of Utilization in NeurosurgeryPublished by S. Karger AG ,2015
- Editorials: BRAIN SCANNING—IS IT BECOMING UNNECESSARILY COMPLICATED?American Journal of Roentgenology, 1971
- Ergebnisse der digitalen Gammaencephalographie bei Hirntumoren; Vergleich von Wismut206, Quecksilber203-Neohydrin und Technetium99mArchiv Fur Psychiatrie Und Nervenkrankheiten, 1967
- Use of Radioactive Diiodofluorescein in the Diagnosis and Localization of Brain TumorsScience, 1948