Clinical effectiveness of the Du Pont aca measurement of creatine kinase MB in serum from patients in a coronary-care unit.
Open Access
- 1 September 1984
- journal article
- research article
- Published by Oxford University Press (OUP) in Clinical Chemistry
- Vol. 30 (9) , 1552-1554
- https://doi.org/10.1093/clinchem/30.9.1552
Abstract
We evaluated the clinical effectiveness of measuring creatine kinase (CK; EC 2.7.3.2) isoenzyme MB and lactate dehydrogenase (LD; EC 1.1.1.27) isoenzymes in diagnosis of acute myocardial infarction. We used an agarose electrophoresis method to measure CK and LD isoenzymes and the Du Pont aca column method to measure CK-MB. Serial blood specimens were drawn from 100 patients consecutively admitted to our Coronary Care Unit. Because of the low diagnostic specificity for CK-MB measurements by both agarose electrophoresis and the discrete-analysis method, as compared with reported values, we re-evaluated our isoenzyme data by using Receiver Operating Characteristic curves. Such analysis of the data established optimal decision levels of greater than or equal to 25 U/L and greater than or equal to 18 U/L plus greater than or equal to 6% of total CK for serum CK-MB measured by the agarose electrophoresis and the aca methods, respectively, and an optimal decision level of greater than or equal to 0.92 for the ratio of LD 1/2 measured after agarose electrophoresis. At these decision levels we obtained a sensitivity of 100%, 100%, and 95% and a specificity of 94%, 92%, and 90% for CK-MB (agarose electrophoresis), CK-MB (aca), and the LD 1/2 ratio, respectively.This publication has 4 references indexed in Scilit:
- Use of receiver operating characteristic curves to evaluate the clinical performance of analytical systems.Clinical Chemistry, 1981
- Comparison of an agarose electrophoresis method and four ion-exchange methods for the measurement of the creatine kinase MB fraction in human serumClinical Biochemistry, 1981
- Optimal diagnosis in acute myocardial infarction. A cost-effectiveness study.Circulation, 1980
- Thin-layer agarose electrophoresis of lactate dehydrogenase isoenzymes in serum: a note on the method of reporting and on the lactate dehydrogenase isoenzyme-1/isoenzyme-2 ratio in acute myocardial infarction.Clinical Chemistry, 1979