A deconvolution scatter correction for a 3-D PET system
- 1 January 1992
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Medical Imaging
- Vol. 11 (4) , 560-569
- https://doi.org/10.1109/42.192692
Abstract
A method to remove the scattered background from a reconstructed image by deconvolution with a point response function which includes the scatter contribution is presented. The amplitude of the scattered response function is obtained by constraining a region of the corrected image to zero average amplitude. This method assumes that the shape of scatter distribution is shift invariant and independent of the shape of the scattering object and the distribution of the positron activity. The validity of these approximations for the QPET geometry was tested using simulations. An average scatter response function for the system was obtained from these simulations and compared with results from measurements. The method was tested using experimental data from an irregularly shaped acrylic phantom. It was simple to implement and resulted in a satisfactory correction of the scattered background for a small-volume system.Keywords
This publication has 20 references indexed in Scilit:
- Cross-plane scattering correction-point source deconvolution in PETIEEE Transactions on Medical Imaging, 1991
- Accurate attenuation correction for a 3D PET systemPhysics in Medicine & Biology, 1991
- Image reconstruction for a 3D PET system using a minimum norm constraintPhysics in Medicine & Biology, 1991
- Three-dimensional image reconstruction from complete projectionsPhysics in Medicine & Biology, 1989
- The effect of collimation on single rates in multi-slice PETIEEE Transactions on Nuclear Science, 1989
- Description and performance of a prototype PET system for small volume imagingNuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1988
- Compton scattering in a large-aperture positron imaging systemIEEE Transactions on Medical Imaging, 1988
- Software correction of scatter coincidence in positron CTEuropean Journal of Nuclear Medicine and Molecular Imaging, 1984
- Correction for Scattered Radiation in a Ring Detector Positron Camera by Integral Transformation of the ProjectionsJournal of Computer Assisted Tomography, 1983
- Noise Identification and Removal in Positron Imaging SystemsIEEE Transactions on Nuclear Science, 1981