The PreOp flexible sigmoidoscopy trainer
- 1 October 2002
- journal article
- research article
- Published by Springer Nature in Surgical Endoscopy
- Vol. 16 (10) , 1459-1463
- https://doi.org/10.1007/s00464-002-9014-4
Abstract
AIM: To demonstrate face and construct validity of a computer based flexible sigmoidoscopy trainer. METHODS: The PreOp (Immersion Medical, USA) system is a virtual reality based flexible sigmoidoscopy simulator. The system records several performance parameters, such as percentage of colonic mucosa visualized, time taken, and pathlength of endoscope travel. Forty-five subjects were divided into three groups: novice (never performed a lower GI endoscopy), intermediate (5–50 examinations), and trained (greater than 200 examinations). After initial familiarization subjects were assessed three times on a case module. Results showed a nonparametric distribution. RESULTS: There was a significant difference between all three groups with respect to percentage of mucosa visualized (novice 71.0 ± 3.7%, intermediate 77.3 ± 5.6%, expert 84.8 ± 4.6%, Kruskal–Wallis p <0.001) and efficiency ratio (%mucosa/time, novice 0.163 ± 0.055, intermediate 0.259 ± 0.07, expert 0.306 ± 0.058, p <0.001). The novice group was also slower and had a lower pathlength of instrument travel compared to the others. CONCLUSION: PreOp virtual reality simulator is a valid discriminator of flexible sigmoidoscopic experience. Its effect on training needs to be explored.Keywords
This publication has 0 references indexed in Scilit: