Design, Construction, and Two Applications for an Automated Flow—Cell Polarization Fluorometer with Digital Read Out: Enzyme-Inhibitor (Antitrypsin) Assay and Antigen—Antibody (Insulin—Insulin Antiserum) Assay

Abstract
We have constructed and automated a flow cell polarization fluorometer and demonstrated two specific clinical applications of fluorescence polarization assay, i.e., the enzyme-inhibitor assay of antitrypsin in serum and the antigen-antibody assay of insulin and its antibody. A signal is displayed directly and is immediately available for chart recording and (or) digital data processing. Since fluorescence polarization offers a number of choices in assay parameters and the use of different fluorescent probes, its development for rapid simultaneous measurements of multiple components in body fluids should be possible.

This publication has 0 references indexed in Scilit: