Effect of Helium in Helium Headspace Carbon Dioxide Cylinders on Packed-Column Supercritical Fluid Chromatography
- 1 March 1996
- journal article
- research article
- Published by American Chemical Society (ACS) in Analytical Chemistry
- Vol. 68 (5) , 894-898
- https://doi.org/10.1021/ac950922y
Abstract
Supercritical fluid chromatography of PAHs was performed with pure carbon dioxide and helium headspace carbon dioxide at various cylinder fill levels. The retention times of the PAHs increased when helium headspace carbon dioxide was used as a carrier fluid relative to pure carbon dioxide. The increased retention times were affected by the level of the liquid phase present in the helium headspace carbon dioxide cylinder. As more liquid phase was removed from the cylinder, the effect of helium on the solvating power of CO2 was reduced because the relative amount of helium dissolved in the liquid phase decreased. Furthermore, the effect of helium headspace carbon dioxide was investigated with methanol-modified carbon dioxide mobile phases for the analysis of steroids. We observed that the relative solubility of helium in carbon dioxide resulted in longer retention times when compared to pure carbon dioxide as the liquid level of carbon dioxide decreased.Keywords
This publication has 5 references indexed in Scilit:
- Effect of Entrained Helium on the Kinetics of Supercritical Fluid Extraction with Carbon DioxideJournal of Chromatographic Science, 1994
- Supercritical fluid chromatography and extractionAnalytical Chemistry, 1992
- Comparison of helium head pressure carbon dioxide and pure carbon dioxide as mobile phases in supercritical fluid chromatographyJournal of Chromatography A, 1990
- Reproducibility of packed-column supercritical-fluid chromatography with helium head-pressure carbon dioxideJournal of Chromatography A, 1989
- Effects of fluid filling techniques on reproducibility in capillary SFCJournal of High Resolution Chromatography, 1987