Importance of Periodic Reeducation of Hospital-Based Nurses in Capillary Blood Glucose Monitoring and an Evaluation of the Usefulness of Reflectance Meters
- 1 October 1989
- journal article
- research article
- Published by SAGE Publications in The Diabetes Educator
- Vol. 15 (5) , 435-439
- https://doi.org/10.1177/014572178901500513
Abstract
Precision and accuracy of capillary blood glucose measure ments performed by general staff nurses were assessed before and after a modest reeducation pro gram both with and without the use of reflectance meters. A total of 380 capillary glucose determinations were performed by nurses and the hospital laboratory on the same samples. Prior to reeducation, nurses using meters were more accurate than those reading visually, but no difference was found in precision. Reeducation improved precision equally in both groups while improving accuracy only in the group using meters. While glucose readings with the use of reflectance meters were statistically more accurate than those read visually, the percentage of readings within 10 % and 20 % of the laboratory readings was the same. Thus, the increased accuracy observed with meters may have little practical significance, suggesting that their use might better be based on other consider ations, such as user confidence, convenience, cost, and quality control. Periodic education is, however, an essential component of any quality assurance program.Keywords
This publication has 13 references indexed in Scilit:
- Implementation of a Successful Bedside Capillary Blood Glucose Monitoring Program by Registered Nurses in an Acute Care FacilityThe Diabetes Educator, 1988
- Bedside capillary glucose measurement by staff nurses in a general hospitalThe American Journal of Medicine, 1986
- Bedside Capillary Glucose Monitoring in the General HospitalDiabetes Care, 1984
- Comparative Study of Home Blood Glucose Monitoring Devices: Visidex, Chemstrip bG, Glucometer, and Accu-Chek bGDiabetes Care, 1983
- Accuracy of Two Systems for Blood Glucose Monitoring without a Meter (Chemstrip/Visidex)Diabetes Care, 1983
- Comparison of Blood Glucose Testing Using Reagent Strips With and Without a Meter (Chemstrips bG and Dextrostix/Dextrometer)Diabetes Care, 1981
- Comparison of Methods for Blood Glucose MonitoringDiabetes Care, 1981
- Comparison of Various Methods for Rapid Glucose EstimationDiabetes Care, 1981
- A Comparison of Accuracy and Estimated Cost of Methods for Home Blood Glucose MonitoringDiabetes Care, 1981
- Self-Monitoring of Blood Glucose Without a MeterDiabetes Care, 1981