Validation of the APTIMA Combo 2 Assay for the Detection of Chlamydia trachomatis and Neisseria gonorrhoeae in SurePath Liquid-Based Pap Test Samples Taken With Different Collection Devices
- 1 September 2009
- journal article
- research article
- Published by Wolters Kluwer Health in Sexually Transmitted Diseases
- Vol. 36 (9) , 581-583
- https://doi.org/10.1097/olq.0b013e3181a4c3fc
Abstract
Mocked samples of Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) diluted in SurePath liquid-based Pap (L-Pap) fluid were detected by the APTIMA Combo 2 assay to end points 10-fold greater than dilutions in specimen transport media. Pooled L-Pap clinical specimens yielded CT-positive results after storage at room temperature for 10 days. Based on an infected patient standard for comparison, cervical swabs, urine, and SurePath L-Pap test samples collected with a SurePath cervical broom or ThinPrep cytobrush from 520 women then tested by APTIMA Combo 2 assay, detected 25 (4.8%) with CT, 5 (1.0%) with (GC), and 3 (0.6%) with both. Percent sensitivities (80-84), specificities (99.8-100), positive (99.5-100) and negative (99.2-99) predictive values of SurePath L-Pap for CT were validated as similar to those reported in a previously published multicenter trial. All values for GC were 100%. One collection device was not significantly better than the other.Keywords
This publication has 7 references indexed in Scilit:
- Abilities of APTIMA, AMPLICOR, and ProbeTec Assays To Detect Chlamydia trachomatis and Neisseria gonorrhoeae in PreservCyt ThinPrep Liquid-Based Pap SamplesJournal of Clinical Microbiology, 2007
- Detection of Chlamydia trachomatis and Neisseria gonorrhoeae Infections in North American Women by Testing SurePath Liquid-Based Pap Specimens in APTIMA AssaysJournal of Clinical Microbiology, 2007
- Recovery of human papillomavirus nucleic acids from liquid-based cytology mediaJournal of Virological Methods, 2006
- Molecular diagnosis of human papillomavirus (HPV) infectionsJournal of Clinical Virology, 2005
- Additional collection devices used in conjunction with the SurePath Liquid-Based Pap Test broom device do not enhance diagnostic utilityBMC Women's Health, 2004
- Comparison of the APTIMA CT and GC Assays with the APTIMA Combo 2 Assay, the Abbott LCx Assay, and Direct Fluorescent-Antibody and Culture Assays for Detection of Chlamydia trachomatis and Neisseria gonorrhoeaeJournal of Clinical Microbiology, 2004
- Human Papillomavirus DNA Testing as an Adjunct to Cytology in Cervical Screening ProgramsArchives of Pathology & Laboratory Medicine, 2003