The Effect of Metal Detector Gates on Implanted Permanent Pacemakers
- 1 October 1988
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 11 (10) , 1386-1397
- https://doi.org/10.1111/j.1540-8159.1988.tb04985.x
Abstract
The effect of metal detector security gates, such as are used in airports, was tested in 103 nonselected pacemaker patients. Various types of single and dual chamber units were examined, using telemetry during the test. Pulse rate and duration were measured immediately before and after the procedure. No ill effect was seen on any of the units tested, pacemaker inhibition was not observed, and programmability was not affected. Metal detector security gates have no effect on implanted permanent pacemakers.Keywords
This publication has 22 references indexed in Scilit:
- Usefulness of the ajmaline test in patients with latent bundle branch blockThe American Journal of Cardiology, 1982
- Intermittent Atrioventricular Block: Procainamide Administration as a Provocative Test*Australian and New Zealand Journal of Medicine, 1978
- A Prospective Study of Sudden Death in “High-Risk” Bundle-Branch BlockNew England Journal of Medicine, 1978
- Electrophysiologic effects of procainamide in subtherapeutic to therapeutic doses on human atrioventricular conduction systemThe American Journal of Cardiology, 1976
- Atrioventricular BlockNew England Journal of Medicine, 1975
- Syncope in Patients with Chronic Bifascicular BlockAnnals of Internal Medicine, 1974
- Electrophysiologic properties of procainamide in manThe American Journal of Cardiology, 1974
- Conduction Disturbances Located within the His BundleCirculation, 1972
- His Bundle Electrogram in Bundle-Branch BlockCirculation, 1972
- Relationship of Right Bundle-Branch Block and Marked Left Axis Deviation (with Left Parietal or Peri-infarction Block) to Complete Heart Block and SyncopeCirculation, 1968