A system for doing low temperature-high pressure single crystal X-ray diffraction with a synchrotron source
- 1 December 1991
- journal article
- research article
- Published by Taylor & Francis in High Pressure Research
- Vol. 6 (6) , 379-388
- https://doi.org/10.1080/08957959208201046
Abstract
A new diamond anvil cell and a helium flow cryostat have been developed for X-ray diffraction on single crystals at low temperatures and high pressures using white radiation of a synchrotron beam. This novel instrument especially enables continuous change of temperature and pressure of the sample without any adjustment of alignment. Automatic search for diffraction peaks can be performed since less than 30 pm eccentricity can be maintained during the rotation of the cell in the cryostat and the rotation of the cryostat on the goniometer head. The minimum temperature reached is 46 K. Measurements of solid 4He at 11.8 GPa are presented which confirm the stability of the hcp phase on this isobar.Keywords
This publication has 8 references indexed in Scilit:
- Theory of high-pressure phases of hydrogenPhysical Review Letters, 1989
- Equation of State of Dense HeliumPhysical Review Letters, 1988
- The membrane diamond anvil cell: A new device for generating continuous pressure and temperature variationsHigh Pressure Research, 1988
- High-Pressure Phase Diagram and Equation of State of Solid Helium from Single-Crystal X-Ray Diffraction to 23.3 GPaPhysical Review Letters, 1988
- Synchrotron X-ray Diffraction Measurements of Single-Crystal Hydrogen to 26.5 GigapascalsScience, 1988
- Three-body exchange interaction in dense heliumPhysical Review Letters, 1987
- Calibration of the ruby pressure gauge to 800 kbar under quasi‐hydrostatic conditionsJournal of Geophysical Research, 1986
- Miniature diamond anvil pressure cell for single crystal x-ray diffraction studiesReview of Scientific Instruments, 1974