Acoustical absorption spectroscopy of liquids between 0.15 and 3000 MHz: II. Ultrasonic pulse transmission methods
- 1 January 1988
- journal article
- Published by IOP Publishing in Journal of Physics E: Scientific Instruments
- Vol. 21 (1) , 98-103
- https://doi.org/10.1088/0022-3735/21/1/018
Abstract
Two methods are described which allow precise measurements of the ultrasonic absorption coefficient of liquids between about 5 and 400 MHz and 1 and 50 MHz, respectively. In both cases RF substitution techniques are applied to determine the attenuation of pulsed signals transmitted through the sample cell. In the former method a cell with variable spacing of the piezoelectric transmitter and receiver crystal is used and the measurements are automatically driven by a process control computer. The latter method is based on a cell of fixed pathlength, using a simple electronic set-up and a manual mode of operation. The construction of the cells is given in detail and the relative merits and limitations of the two methods are briefly discussed. Examples of acoustic absorption spectra are presented to demonstrate the accuracy of measurements.Keywords
This publication has 3 references indexed in Scilit:
- Below cut-off piston attenuator as a calculable signal vernier for microwaves up to 15 GHzJournal of Physics E: Scientific Instruments, 1986
- Ultrasonic absorption and velocity measurements at mW peak power level in the range 50-500 MHz with variable pathlength cellJournal of Physics E: Scientific Instruments, 1985
- Ultrasonic Measurements with Milliliter Liquid Samples in the 0.5–100 MHz RangeReview of Scientific Instruments, 1973