Prosthetic Valves for the Small Aortic Root
- 1 March 1994
- journal article
- Published by Hindawi Limited in Journal of Cardiac Surgery
- Vol. 9 (s2) , 154-157
- https://doi.org/10.1111/j.1540-8191.1994.tb00916.x
Abstract
Mechanical aortic valves (AVs) are frequently implanted in small (19 and 21 mm) aortic roots because bioprosthetic valves have unacceptably high gradients and many surgeons do not implant allograft valves. Three mechanical valves in common use today in the United States are the Starr-Edwards (SE), St. Jude Medical (SJ), and the Medtronic-Hall (MH). Clinical hemodynamic studies reveal that the 21-mm SE valve (size 8A) has peak systolic gradients of 13 to 58 mmHg (N = 6) with a calculated effective orifice of 0.7 to 1.4 cm2. The 19-mm SJ valve has a gradient at peak pressure of 17 mmHg and a mean gradient of 22 mmHg (N = 6) with respective exercise gradients of 32 and 38 mmHg (N = 5). For the 21-mm SJ valve the mean gradient was 5.2 +/- 5.3 (N +/- 12) and the gradient at peak pressure was 6.0 mmHg (N = 15). The 21-mm MH valve had resting gradients at peak pressure of 10.5 (N = 3) and 12.4 mmHg (N = 9) and exercise gradients of 15.8 mmHg (N = 9). Six months after AV replacement with small SJ or MH (N = 14) or large (23 mm or greater) (N = 83) valves, cardiac output was 4.7 versus 6.4 L/min (p < 0.03), percent reduction in left ventricular mass index (LVMI) was -8% versus -21% (p < 0.01), exercise duration was 370 versus 555 seconds, and congestive heart failure (CHF) class was 1.9 versus 1.1 (p < 0.0001). Change in LVMI and valve size were the only independent predictors of CHF class.(ABSTRACT TRUNCATED AT 250 WORDS)Keywords
This publication has 10 references indexed in Scilit:
- Effect of stenosis geometry on the Doppler-catheter gradient relation in vitro: A manifestation of pressure recoveryJournal of the American College of Cardiology, 1993
- Small aortic valve prostheses; inferior clinical and echocardiographic resultsJournal of the American College of Cardiology, 1991
- Continuous wave Doppler echocardiographic measurement of prosthetic valve gradients. A simultaneous Doppler-catheter correlative study.Circulation, 1989
- Continuous-wave Doppler velocities and gradients across fixed tunnel obstructions: studies in vitro and in vivo.Circulation, 1987
- Valve Replacement in Narrow Aortic Roots: Serial Hemodynamics and Long-Term Clinical OutcomeThe Annals of Thoracic Surgery, 1986
- Flow characteristics of four commonly used mechanical heart valvesThe American Journal of Cardiology, 1986
- Primary clinical experience with the Hall-Kaster valve in the aortic position: results at 3 months including hemodynamic studies.Circulation, 1979
- Intraoperative and Postoperative Hemodynamic Studies in Patients Undergoing Aortic Valve Replacement with the Hall-Kaster Cardiac Disc Valve ProsthesisThe Thoracic and Cardiovascular Surgeon, 1979
- Hemodynamic Evaluation of Lillehei-Kaster and Starr-Edwards ProsthesesThe Annals of Thoracic Surgery, 1978
- The problem of valve prosthesis-patient mismatch.Circulation, 1978