Verification of a pencil beam based treatment planning system: output factors for open photon beams shaped with MLC or blocks
- 17 August 1999
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 44 (9) , N201-N207
- https://doi.org/10.1088/0031-9155/44/9/402
Abstract
The accuracy of monitor unit calculations from a pencil beam based, three-dimensional treatment planning system (3D TPS) has been evaluated for open irregularly shaped photon fields. The dose per monitor unit was measured in water and in air for x-ray beam qualities from 6 to 15 MV. The fields were shaped either with a multileaf collimator (MLC) or with customized alloy blocks. Calculations from the 3D TPS were compared with measurements. The agreement between calculated and measured dose per monitor unit depended on field size and the amount of blocking and was within 3% for the MLC-shaped fields. The deviation could be traced to limitations in head scatter modelling for the MLC. For fields shaped with alloy blocks, the dose per monitor unit was calculated to be within 1.6% of measured values for all fields studied. The measured and calculated relative phantom scatter for fields with the same equivalent field size were identical for MLC and alloy shaped fields. These results indicate that the accuracy in the TPS calculations for open irregular fields, shaped with MLC or blocks, is satisfactory for clinical situations.Keywords
This publication has 18 references indexed in Scilit:
- A comparison of multileaf-collimator and alloy-block field shapingInternational Journal of Radiation Oncology*Biology*Physics, 1998
- Effects of treatment setup variation on beam's eye view dosimetry for radiation therapy using the multileaf collimator vs. the cerrobend blockInternational Journal of Radiation Oncology*Biology*Physics, 1995
- Modeling transmission and scatter for photon beam attenuatorsMedical Physics, 1995
- The acceptability of a multileaf collimator as a replacement for conventional blocksRadiotherapy and Oncology, 1995
- Collimator scatter in photon therapy beamsMedical Physics, 1995
- Analytic modeling of photon scatter from flattening filters in photon therapy beamsMedical Physics, 1994
- Clinical dosimetry for implementation of a multileaf collimatorMedical Physics, 1992
- Application of the convolution method for calculation of output factors for therapy photon beamsMedical Physics, 1992
- A pencil beam model for photon dose calculationMedical Physics, 1992
- Tissue standard ratios for irregularly shaped radiotherapy fieldsThe British Journal of Radiology, 1990