Artefacts in multi-echoT2imaging for high-precision gel dosimetry: II. Analysis ofB1-field inhomogeneity
- 26 June 2000
- journal article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 45 (7) , 1825-1839
- https://doi.org/10.1088/0031-9155/45/7/308
Abstract
In BANG gel dosimetry, the spin-spin relaxation rate, R2 = 1/T2, is related to radiation dose that has been delivered to a gel phantom. R2 is calculated by fitting the pixel intensities of a set of differently T2-weighted base images. The accuracy that is aimed for in this quantitative MR application is about 5% relative to the maximum dose. In a conventional imaging MR scanner, however, several imaging artefacts may perturb the final dose map. These deviations manifest themselves as either a deformation of the dose map or an inaccuracy of the dose pixel value. Inaccuracies in the dose maps are caused by both spatial and temporal deviations in signal intensities during scanning. This study deals with B1-field inhomogeneities as a source of dose inaccuracy. First, the influence of B1-field inhomogeneities on slice profiles is investigated using a thin-slice phantom. Secondly, a FLASH sequence is used to map the B1-field by assessing the effective flip angle in each voxel of a homogeneous phantom. In addition, both experiments and computer simulations revealed the effects of B1 field inhomogeneities on the measured R2. This work offers a method to correct R2 maps for B1 -field inhomogeneities.Keywords
This publication has 39 references indexed in Scilit:
- Validation of MR-based polymer gel dosimetry as a preclinical three-dimensional verification tool in conformal radiotherapyMagnetic Resonance in Medicine, 2000
- Three-dimensional dosimetry using polymer gel and magnetic resonance imaging applied to the verification of conformal radiation therapy in head-and-neck cancerRadiotherapy and Oncology, 1998
- Distributed large-scale simulation of magnetic resonance imaging.Magnetic Resonance Materials in Physics, Biology and Medicine, 1997
- A Method for Accurate Calculation ofB1Fields in Three Dimensions. Effects of Shield Geometry on Field Strength and Homogeneity in the Birdcage CoilJournal of Magnetic Resonance, 1997
- Precision, accuracy, and image plane uniformity in NMR relaxation time imaging on a 1.5 T whole-body MR imaging systemMagnetic Resonance Imaging, 1994
- Effects of RF amplifier distortion on selective excitation and their correction by prewarpingMagnetic Resonance in Medicine, 1992
- A slice geometry phantom for cross sectional tomographic imagersMedical Physics, 1989
- Errors in T2 estimation using multislice multiple‐echo imagingMagnetic Resonance in Medicine, 1987
- Test Method for MR Image Slice ProfileJournal of Computer Assisted Tomography, 1985
- RF magnetic field penetration, phase shift and power dissipation in biological tissue: implications for NMR imagingPhysics in Medicine & Biology, 1978