Leakage of fluid around high‐volume, low‐pressure cuffs
- 1 January 2001
- journal article
- research article
- Published by Wiley in Anaesthesia
- Vol. 56 (1) , 38-42
- https://doi.org/10.1046/j.1365-2044.2001.01718.x
Abstract
We studied the ability of high-volume, low-pressure tracheal tube cuffs (Portex Soft Seal®, Portex Profile®, Mallinckrodt Lo-Contour® and Mallinckrodt Hi-Lo® tubes) to prevent leakage of fluid into the airway, in a model trachea and lung. Five tubes (7.0 and 8.0 mm internal diameter) of each type were used. Each tube was inserted into the model trachea and the cuff inflated until the intracuff pressure reached either 20, 30 or 40 cmH2O. The model lung was ventilated with a tidal volume of 700 ml and respiratory rate of 14 breath.min−1 at a compliance of 20 cmH2O. Ten millilitres of 0.01% methylene blue solution were infused over the cuff and the volume of fluid leaking past the cuff over 5 min was measured. The leak volume for the Soft Seal tube was less than that for the Profile or Lo-Contour tubes at all intracuff pressures (all p < 0.05). Compared with the Hi-Lo tube, the volume leaking past the cuff for the Soft Seal tube was greater at an intracuff pressure of 20 cmH2O (p < 0.05), whereas there was no significant difference between these two tubes at an intracuff pressure of 30 or 40 cmH2O. We conclude that the cuff of the Portex Soft Seal tube prevented leakage of fluid significantly more than that of the Portex Profile or Mallinckrodt Lo-Contour tubes, and to a similar degree to that of the Mallinckrodt Hi-Lo tube.Keywords
This publication has 14 references indexed in Scilit:
- Improving the shape and compliance characteristics of a high-volume, low-pressure cuff improves tracheal seal.British Journal of Anaesthesia, 1999
- Leakage of fluid around low‐pressure tracheal tube cuffsAnaesthesia, 1997
- GASTRODUODENAL DYSFUNCTION AS A CAUSE OF GASTRIC BACTERIAL OVERGROWTH IN PATIENTS UNDERGOING MECHANICAL VENTILATION OF THE LUNGSBritish Journal of Anaesthesia, 1992
- Pulmonary Aspiration of Gastric Contents in Patients Receiving Mechanical Ventilation: The Effect of Body PositionAnnals of Internal Medicine, 1992
- Airway Considerations in the Management of Patients Requiring Long-Term Endotracheal IntubationAnesthesia & Analgesia, 1992
- Aspiration beyond endotracheal cuffsCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1986
- Dimensions of the growing trachea related to age and genderAmerican Journal of Roentgenology, 1986
- Dimensions of the normal human tracheaAmerican Journal of Roentgenology, 1984
- Endotracheal cuff pressure and tracheal mucosal blood flow: endoscopic study of effects of four large volume cuffs.BMJ, 1984
- Failure of a High-Compliance Low-pressure Cuff to Prevent AspirationAnesthesiology, 1975