A High Resolution Echelle Monochromator and Its Application to Ion Temperature Measurement of He Plasma
- 1 October 1966
- journal article
- Published by IOP Publishing in Japanese Journal of Applied Physics
- Vol. 5 (10)
- https://doi.org/10.1143/jjap.5.970
Abstract
For the measurement of the profile of a spectral line, a high dispersive monochromator with an Echelle grating and 12 channels of photoelectric devices using thin glass sheets to separate the line profile was constructed. The measured resolving power is about 2×105 and the linear dispersion at the image plane is 1.2 mm/Å at 5000Å. This instrument was used to measure the ion temperature of a BSG device by employing the 4686Å line of He II.Keywords
This publication has 8 references indexed in Scilit:
- A Multichannel Fabry-Perot InterferometerApplied Optics, 1965
- Project MACETIE (Multistage adiabatic compression and equi-temperature irreversible expansion)Nuclear Fusion, 1965
- Measurements of Ion Teraperature and Ion Collective Motion in an Energetic Arc.Kakuyūgō kenkyū, 1965
- Ultra High Speed Spectrometer for Measurement of Spectral Line ProfileJapanese Journal of Applied Physics, 1964
- Doppler Temperatures in the C StellaratorPhysics of Fluids, 1964
- Electronic-Recording, Time-Resolving SpectrometerJournal of the Optical Society of America, 1958
- The Luminosity of Spectrometers with Prisms, Gratings, or Fabry-Perot EtalonsJournal of the Optical Society of America, 1954