Response of Liquid-Membrane Calcium-Selective Electrodes to Calcium Ion Activity Steps
- 1 March 1991
- journal article
- research article
- Published by Taylor & Francis in Analytical Letters
- Vol. 24 (3) , 505-518
- https://doi.org/10.1080/00032719108052921
Abstract
The results of fast Ca2+ activity step experiments using ionophore-and ion exchanger-based Ca2+-selective liquid membrane electrodes are presented. The recorded curves indicate that there is little dependence of electrode response on activity range or active electrode material. The response is shown to be dependent on the age of the ion-selective electrode (ISE) membrane. Impedance experiments performed show large fluctuations in the membrane bulk resistance over time. These results suggest that the membrane rapidly changes structure over a short period of time, which results in a slowing of the ISE response.Keywords
This publication has 26 references indexed in Scilit:
- Anomalies In Theories of Liquid Membrane Time Responses to Activity StepsAnalytical Letters, 1989
- Switched wall jet for dynamic response measurementsAnalytical Chemistry, 1987
- Evaluation of the optimum composition of neutral-carrier membrane electrodes with incorporated cation-exchanger sitesAnalytica Chimica Acta, 1984
- Evaluation of calcium ion-selective electrodes based on di(n-alkylphenyl)-phosphate sensors and their calibration with ion buffersThe Analyst, 1979
- Response time studies on neutral carrier ion-selective membrane electrodesAnalytical Chemistry, 1978
- Calcium ion-selective electrodes based on calcium bis[di(p-1,1,3,3-tetramethylbutylphenyl)-phosphate] sensor and trialkyl phosphate mediatorsThe Analyst, 1978
- A calcium-sensitive microelectrode suitable for intracellular measurement of calcium(II) activityAnalytica Chimica Acta, 1976
- Factors affecting the response time of a calcium selective liquid membrane electrodeAnalytical Chemistry, 1974
- Selectrode—the universal ion-selective electrode: Part VI. The calcium(II) selectrode employing a new ion exchanger in a nonporous membrane and a solid-state reference systemAnalytica Chimica Acta, 1973