An Experimental Study on Transpedicular Screw Fixation in Relation to Osteoporosis of the Lumbar Spine

Abstract
In order to elucidate the relationship between the severity of osteoporosis and the fixation strength of a pedicle screw, screw pull-out tests were performed using cadaveric lumbar vertebrae. The severity of osteoporosis was evaluated by the Jikei osteoporosis grading scale (Jikei method), bone mineral density, and microdensitometry. When a 7.0-mm screw was used, the pull-out force of the screw was 1,056.4 N in the normal group (as determined by the Jikei method), while it was 495.6 N in the Grade I osteoporosis and 269.5 N in the Grade II osteoporosis groups, respectively. There were also positive correlations between the pull-out force and bone mineral density and each parameter of the microdensitometry method. When bone cement was used in an osteoporotic vertebra, twofold stronger pull-out force was obtained in comparison to that obtained without bone cement.