Characterization of a mini membrane inlet mass spectrometer for on‐site detection of contaminants in both aqueous and liquid organic samples
- 3 April 2006
- journal article
- research article
- Published by Wiley in Rapid Communications in Mass Spectrometry
- Vol. 20 (9) , 1441-1446
- https://doi.org/10.1002/rcm.2466
Abstract
A mini membrane inlet mass spectrometer (mini‐MIMS) of a total weight of 12 kg was constructed using a miniature Multipole mass spectrometer, a small vacuum system and a flexible flat sheet membrane inlet, where the exposed membrane area can be changed by a factor of 80. The variable membrane area together with the possibility of operating the Multipole at pressures up to 1 × 10−3 Torr made it possible to test the system with three microporous membranes (cellulose, polyether sulfone and polypropylene) normally not compatible with standard electron ionization MIMS systems and a standard non‐porous polydimethylsiloxane membrane. We found that the hydrophilic cellulose and polyether sulfone membranes had selectivity characteristics opposite to those of the standard silicone membrane. They demonstrated preferential detection of hydrophilic compounds in hydrophobic organic solvents, whereas the silicone membrane preferentially detects hydrophobic organic compounds in aqueous solution. Using the cellulose membrane, organic contaminants and water could be detected in organic solvents at 10–100 ppm levels by weight, the relative high detection limits primarily caused by interference from a high chemical background from the solvent. When being used with the standard silicone membrane the mini‐MIMS behaved just like most standard MIMS systems with detection limits of volatile organic compounds in water at concentrations just below 1 ppm. The hydrophobic microporous polypropylene membrane was not found to be useful with the mini‐MIMS. Copyright © 2006 John Wiley & Sons, Ltd.Keywords
This publication has 28 references indexed in Scilit:
- NEREUS/Kemonaut, a mobile autonomous underwater mass spectrometerTrAC Trends in Analytical Chemistry, 2004
- Comparison of analytical performances of a micro‐array quadrupole instrument and a conventional quadrupole mass spectrometer equipped with membrane inletsRapid Communications in Mass Spectrometry, 2003
- Mixture Analysis Using Membrane Introduction Mass Spectrometry and Multivariate CalibrationAnalytical Chemistry, 1997
- High-pressure effects in miniature arrays of quadrupole analyzers for residual gas analysis from 10−9 to 10−2 TorrJournal of Vacuum Science & Technology A, 1996
- ADVANCES IN MEMBRANE INLET MASS SPECTROMETRY (MIMS)Reviews in Analytical Chemistry, 1996
- Trace level analysis of volatile and semi-volatile organic compounds in water using a membrane/jet separator interfaced to an ion trap mass spectrometerAnalytica Chimica Acta, 1995
- On-Site Environmental Analysis by Membrane Inlet Mass SpectrometryAnalytical Chemistry, 1995
- Membrane inlet manometry for the measurement of water activity in aqueous and organic solutionsAnalytica Chimica Acta, 1993
- Microporous membrane introduction mass spectrometry with solvent chemical ionization and glow discharge for the direct detection of volatile organic compounds in aqueous solutionAnalytica Chimica Acta, 1992
- Continuous measurement of water in organic solution by membrane‐inlet mass spectrometryRapid Communications in Mass Spectrometry, 1991