Quantitative comparison of wear debris from UHMWPE that has and has not been sterilised by gamma irradiation
- 1 March 1998
- journal article
- research article
- Published by British Editorial Society of Bone & Joint Surgery in The Journal of Bone and Joint Surgery. British volume
- Vol. 80 (2) , 340-344
- https://doi.org/10.1302/0301-620x.80b2.8234
Abstract
Ultra-high-molecular-weight polyethylene (UHMWPE) components for total joint replacement generate wear particles which cause adverse biological tissue reactions leading to osteolysis and loosening. Sterilisation of UHMWPE components by gamma irradiation in air causes chain scissions which initiate a long-term oxidative process that degrades the chemical and mechanical properties of the polyethylene. Using a tri-pin-on-disc tribometer we studied the effect of ageing for ten years after gamma irradiation in air on the volumetric wear, particle size distribution and the number of particles produced by UHMWPE when sliding against a stainless-steel counterface. The aged and irradiated material produced six times more volumetric wear and 34 times more wear particles per unit load per unit sliding distance than non-sterilised UHMWPE. Our findings indicate that oxidative degradation of polyethylene after gamma irradiation in air with ageing produces more wear.Keywords
This publication has 0 references indexed in Scilit: