Performance of a Dual-Range Cloud Nucleus Counter
Open Access
- 1 September 1978
- journal article
- Published by American Meteorological Society in Journal of Applied Meteorology
- Vol. 17 (9) , 1286-1297
- https://doi.org/10.1175/1520-0450(1978)017<1286:poadrc>2.0.co;2
Abstract
An instrument for measuring the concentrations of cloud condensation nuclei as a function of nucleus critical supersaturation (Se) is described. In the range 0.1% ≤ Se ≤ 1%, the instrument is operated as a vertical flow thermal diffusion chamber. In the range 0.016% ≤ Se ≤ 0.1%, the chamber is operated at 100% humidity, and the equilibrium radii of the nuclei are measured with an optical counter. This paper presents a large body of data taken in the laboratory to evaluate the performance of the instrument. The data indicates that the instrument is internally consistent over a wide range of operating conditions. Abstract An instrument for measuring the concentrations of cloud condensation nuclei as a function of nucleus critical supersaturation (Se) is described. In the range 0.1% ≤ Se ≤ 1%, the instrument is operated as a vertical flow thermal diffusion chamber. In the range 0.016% ≤ Se ≤ 0.1%, the chamber is operated at 100% humidity, and the equilibrium radii of the nuclei are measured with an optical counter. This paper presents a large body of data taken in the laboratory to evaluate the performance of the instrument. The data indicates that the instrument is internally consistent over a wide range of operating conditions.Keywords