Bias and variability in blood pressure measurement with ambulatory recorders.
- 1 October 1993
- journal article
- clinical trial
- Published by Wolters Kluwer Health in Hypertension
- Vol. 22 (4) , 591-598
- https://doi.org/10.1161/01.hyp.22.4.591
Abstract
This study sought to determine whether patient characteristics such as age, sex, blood pressure, and pulse pressure differently affect the accuracy of an oscillometric (SpaceLabs 90207) and a microphonic (TM2420 version 7) blood pressure monitor. Blood pressure recorded by two oscillometric and two microphonic ambulatory monitors was compared with simultaneous readings by two pairs of trained, blinded observers using random-zero sphygmomanometry. One hundred and eighteen subjects (53 men and 65 women, aged 17 to 94 years; systolic pressure, 89 to 211 mm Hg; diastolic, 44 to 116 mm Hg) were studied. There were no significant differences within each observer pair or between the two observer pairs as well as no correlation between interobserver differences and patient characteristics. The differences between the monitor and trained observers' readings were 2.8 +/- 9.9 mm Hg systolic and 3.9 +/- 6.8 mm Hg diastolic for the SpaceLabs and 5.0 +/- 5.2 mm Hg systolic and 3.4 +/- 6.1 mm Hg diastolic for the TM2420. Patient characteristics that predicted measurement error were defined by multiple regression. For oscillometry, systolic measurement error was highly correlated with systolic pressure, pulse pressure, and subject age. The diastolic error was significantly correlated with pulse pressure, diastolic pressure, and subject sex. For the oscillometric monitor, patient characteristics accounted for 36.6% of the variation of the systolic error and 34.7% of the variation of the diastolic error. For the microphonic monitor, only age correlated with diastolic error, and no significant correlations were seen with systolic error. Patient characteristics accounted for only 1.2% of the systolic and 8.9% of the diastolic error.(ABSTRACT TRUNCATED AT 250 WORDS)Keywords
This publication has 21 references indexed in Scilit:
- Pseudohypertension: a new assessmentJournal Of Hypertension, 1993
- Accuracy of the SpaceLabs 90207 determined by the British Hypertension Society ProtocolJournal Of Hypertension, 1991
- Bench and ambulatory field evaluation of the A & D TM-2420 automated sphygmomanometerJournal Of Hypertension, 1990
- Assessment of four ambulatory blood pressure monitors and measurements by clinicians versus intraarterial blood pressure at rest and during exerciseThe American Journal of Cardiology, 1990
- ACCURACY OF THE TAKEDA TM‐2420 AMBULATORY BLOOD PRESSURE MONITORClinical and Experimental Pharmacology and Physiology, 1989
- Validation of portable noninvasive blood pressure monitoring devices: Comparisons with intra-arterial and sphygmomanometer measurementsAmerican Heart Journal, 1988
- A comparison of the random-zero and standard mercury sphygmomanometers.Hypertension, 1988
- A new compact 24-hour indirect blood-pressure recorder and its clinical application.Japanese Heart Journal, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Evaluation of performance of selected devices for measuring blood pressureThe American Journal of Cardiology, 1973