A silicon diode in a thimble-type mount for measurement of diagnostic X-ray spectra
- 1 November 1990
- journal article
- research article
- Published by IOP Publishing in Physics in Medicine & Biology
- Vol. 35 (11) , 1505-1517
- https://doi.org/10.1088/0031-9155/35/11/006
Abstract
A silicon detector which can be used like a thimble chamber has been constructed. The silicon diode chip used here is very small (1.7*1.7*0.1 mm3), and the construction materials of the detector mount are thin and have low atomic numbers. Thus, the characteristic X-rays from the mount give no distortion on the obtained spectra, and the photons incident on the back surface of the silicon chip can be measured. Furthermore, the perturbation of the X-ray field is small. The energy resolution of the detector is 2.0 keV (full width at half maximum) for 59.5 keV photons at room temperature. Bremsstrahlung spectra (60-100 kV, 1-2 mA) have been measured without using a pinhole collimator: the obtained spectra are in good agreement with those obtained with a Ge detector. The spectrum of the radiation in a water phantom irradiated with 90 kV X-rays has also been measured by inserting the detector in the phantom.This publication has 15 references indexed in Scilit:
- Photoelectron escape from a silicon X-ray detector: Monte Carlo studyPhysics in Medicine & Biology, 1990
- Monte Carlo simulation studies of detectors used in the measurement of diagnostic x-ray spectraMedical Physics, 1980
- Stripping of X-ray bremsstrahlung spectra up to 300 kVp on a desk type computerPhysics in Medicine & Biology, 1979
- A comparison of X-ray spectra and outputs from molybdenum and tungsten targetsThe British Journal of Radiology, 1975
- Contribution au calcul des conditions optimales d'irradiation et de mesure de la radioactivité en analyse par activationNuclear Instruments and Methods, 1971
- Spectrometry of low-energy γ- and X-rays with Ge(Li) detectorsNuclear Instruments and Methods, 1967
- Spectral Fluence of Scattered Radiation in a Water Medium Irrradiated with Diagnostic X-RaysRadiation Research, 1967
- Spectral Flux Density of Scattered and Primary Radiation Generated at 250 KvRadiation Research, 1961
- Scattered radiation in a water phantom irradiated by roentgen photons between 50 and 250 keVActa Radiologica, 1960
- Measurement of the Spectral Distribution of Scattered 400 kVp X Rays in a Water PhantomThe British Journal of Radiology, 1957