Microcomputer controlled integrator for Calvet-type twin calorimeters
- 1 May 1980
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 51 (5) , 675-677
- https://doi.org/10.1063/1.1136280
Abstract
An improved integration method has been developed for use with Calvet‐type differential calorimeters. The new integrator is based on a voltage‐to‐frequency converter combined with a digital system interfaced to a microcomputer. The microcomputer provides programmed control of the integrator and, with little or no operator intervention, calculates the total heat effect making corrections for background and baseline drift. With this method a calorimetric precision of the order of ±0.1% has been achieved in experiments carried out at 651°C. This is about one order of magnitude of improvement over graphical methods and about a factor of 3 better than the analog integration techniques which we have used previously.Keywords
This publication has 3 references indexed in Scilit:
- Twin Liquid Metal Solution CalorimeterReview of Scientific Instruments, 1966
- Electronic Integrator for Use with a Differential CalorimeterReview of Scientific Instruments, 1963
- A NEW TWIN HIGH-TEMPERATURE REACTION CALORIMETER. THE HEATS OF MIXING IN LIQUID SODIUM-POTASSIUM NITRATESThe Journal of Physical Chemistry, 1960