Cardiac pacemakers. Does radiation therapy affect performance?
- 6 October 1975
- journal article
- Published by American Medical Association (AMA)
- Vol. 234 (1) , 72-73
- https://doi.org/10.1001/jama.234.1.72
Abstract
Representative deman cardiac pacemakers were tested with several common radiation therapy devices and ancillary equipment. The experimental set-up essentially duplicated the dosimetric and electroconductive environment of the radiation therapy patient bearing a pacemaker. Observations of pulse width and interval confirmed that pacemaker output was generally not appreciably affected by the simulated treatments. Occasionally, changes were caused by ancillary equipment and start-up of treatment. We conclude that although interference hazards are sometimes seen, radiation therapy is not contraindicated in patients wearing pacemakers.Keywords
This publication has 0 references indexed in Scilit: