Occurrence, extent, and implications of pressure waves during excimer laser ablation of normal arterial wall and atherosclerotic plaque
- 1 January 1993
- journal article
- research article
- Published by Wiley in Lasers in Surgery and Medicine
- Vol. 13 (3) , 263-270
- https://doi.org/10.1002/lsm.1900130302
Abstract
Ablation of atherosclerotic plaque and normal arterial wall was performed using a Xenon‐Chloride Excimer laser with a wavelength of 308 nm and a pulse duration of 115 ns. The light was transmitted via a 600 μm bare fibre and adjusted to an energy density of 3.5J/cm2. The acoustic signals generated by the laser pulse were measured with two types of hydrophones consisting of polyvinylidenefluoride with active diameters of 0.3 mm and 0.5 mm and recorded on a dual channel digital storage oscilloscope using either a 0.5 m coaxial cable or a broadband fibre‐optic transmission system. Tissue was retrieved from nine cadaver human aortas and macroscopically classified as either normal or calcified atherosclerotic plaque. Histological analysis (Haematoxylin eosin, elastica van Gieson, and immunohistochemical staining) was carried out after the experiments to verify the macroscopic diagnosis and to correlate the acoustic responses with the tissue characteristics. For normal arterial wall, maximum peak pressure was 1.28 MPa ± 0.85 MPa, rise time 163 ns ± 43 ns, and pressure increase 8,2k Pa ± 5,4k Pa/ns. For calcified, atheromatous segments, a maximum peak pressure of 2,02 MPa ± 1,16 MPa, a rise time of 69,9 ns ± 25,8 ns, and a pressure increase of 32,3 kPa ± 21,3 kPa/ns was found. Statistical analysis showed a significant shorter rise time (P < 0.0001) and a higher pressure increase (P < 0.0001) for calcified tissue in comparison to normal arterial wall, whereas maximum pressures alone did not allow a differentiation of tissue characteristics. Several hundred kPa are generated during Excimer laser ablation. The results suggest that focal tissue fragmentation is one mechanism of plaque ablation. A differentiation of tissue characteristics is possible by analysis of rise time and pressure increase, potentially providing the possibility of acoustic ablation control.Keywords
This publication has 37 references indexed in Scilit:
- Analysis of the acoustic response of vascular tissue irradiated by an ultraviolet laser pulseJournal of Applied Physics, 1991
- Recanalization of chronic total coronary arterial occlusions by percutaneous excimer-laser and laser-assisted angioplastyThe American Journal of Cardiology, 1990
- Percutaneous excimer laser coronary angioplastyThe American Journal of Cardiology, 1990
- Initial angiographic results in ablation of atherosclerotic plaque by percutaneous coronary excimer laser angioplasty without subsequent balloon dilatationThe American Journal of Cardiology, 1989
- Percutaneous excimer laser angioplasty of aortocoronary saphenous vein graftsJournal of the American College of Cardiology, 1989
- Balloon angioplasty in acute and chronic coronary artery diseasePublished by American Medical Association (AMA) ,1989
- Reduction of laser-induced pathologic tissue injury using pulsed energy deliveryThe American Journal of Cardiology, 1985
- Restenosis after percutaneous transluminal coronary angioplasty (PTCA): A report from the PTCA registry of the national heart, lung, and blood instituteThe American Journal of Cardiology, 1984
- Long-term efficacy of percutaneous transluminal coronary angioplasty (PTCA): Report from the national heart, lung, and blood institute PTCA registryThe American Journal of Cardiology, 1984
- Ablative photodecomposition: action of far-ultraviolet (193 nm) laser radiation on poly(ethylene terephthalate) filmsJournal of the American Chemical Society, 1982