Pacing Lead Survival: Performance of Different Models
- 1 November 1992
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 15 (11) , 1991-1995
- https://doi.org/10.1111/j.1540-8159.1992.tb03008.x
Abstract
Recent reports have shown poor survival of some leads currently in use. Long-term survival analysis of 2,444 leads (1,059 atrial and 1,385 ventricular) implanted in this institution since January 1980, and having at least 1 month of follow-up was performed. The survival of 123 different models was compared with the average survival of all the leads implanted in the corresponding chamber. Failure was defined as inactivation of the lead (electrical abandonment, explant, or cap) due to insulator and/or conductor fracture. The mean follow-up was 33 +/- 32 months. The cumulative survival for different atrial lead models was consistent with the average performance in the atrium. No atrial lead showed better or worse survival compared to the others. In the ventricular group, the Medtronic 4012 lead showed statistically significant poorer survival (P = 0.01) compared with the average survival of the ventricular leads. The Cardiac Pacemakers, Inc. (CPI) 4010 lead showed a nonsignificant (P = 0.12) worse performance than the average for ventricular leads. (1) The Medtronic 4012 had a significantly poorer performance than the rest of the leads. A trend in similar direction was found for the CPI 4010, also in the ventricular group; (2) Atrial lead models showed a stable survival; and (3) Frequent follow-up is required for some leads, especially in pacemaker dependent patients.Keywords
This publication has 11 references indexed in Scilit:
- Presalvage prostate‐specific antigen (PSA) and PSA doubling time as predictors of biochemical failure of salvage cryotherapy in patients with locally recurrent prostate cancer after radiotherapyCancer, 2006
- A Multicenter Experience with a Bipolar Tined Polyurethane Ventricular LeadPacing and Clinical Electrophysiology, 1992
- Long‐Term Functional Integrity of Atrial LeadsPacing and Clinical Electrophysiology, 1991
- Appraisal of Pacing Lead Performance Using Transtelephonic Follow‐Up DataPacing and Clinical Electrophysiology, 1990
- Survival of Implantable Pacemaker LeadsPacing and Clinical Electrophysiology, 1990
- Atrial pacemaker leads comparedCleveland Clinic Journal of Medicine, 1990
- Rate Responsive Cardiac Pacing‐A PerspectivePacing and Clinical Electrophysiology, 1989
- Simultaneous Insulation Deterioration Associated with Side‐by‐Side Subclavian Placement of Two Polyurethane LeadsPacing and Clinical Electrophysiology, 1988
- 352. Note: Conservatism of the Approximation Σ(O - E) 2 /E in the Logrank Test for Survival Data or Tumor Incidence DataPublished by JSTOR ,1973
- Nonparametric Estimation from Incomplete ObservationsJournal of the American Statistical Association, 1958