Neonatal hygroscopic condenser humidifier

Abstract
A hygroscopic condenser humidifier was developed for neonates on mechanical ventilation and was evaluated by laboratory tests and clinically. Humidification provided by the unit was measured in the 10− to 50-ml tidal-volume range at ambient temperatures of 24°C and 38°C. The effect of a leaking patient connection on device performance was investigated. Leakage rates were measured routinely in a neonatal ICU and surgery to determine the clinical significance. In the entire tidal volume and temperature range, the unit provided an inspiratory water content in excess of 30 g/m3 when the leak fraction (volume leaked/volume delivered at Y-piece) was less than 15%. This was found in three out of four cases. In about one out of ten cases, the leak exceeded 30%, which invariably led to corrective action, such as repositioning or changing the endotracheal tube. However, even at a 30% leak, a water content of about 26 g/m3 was still available for humidifying the inspired gas, which corresponds to normal physiologic conditions found in the trachea for nasal breathing of room air.