INRs and point of care testing
- 3 July 2003
- Vol. 327 (7405) , 5-6
- https://doi.org/10.1136/bmj.327.7405.5
Abstract
Despite limitations, overall evidence shows that point of care testing is reliable The development and adoption of instruments for determining the international normalised ratio (INR) at the point of care for monitoring treatment with coumarin anticoagulants has added another variable to the question of the precision of INR measurements. In this issue Poller et al (p 30) show discrepancies between results when two such instruments are used and the best estimate of the true INR.1 These observations, however, should be seen in the context of the already recognised limitations of the INR system and the potential for using point of care testing to deliver safe and effective anticoagulant therapy efficiently. Coumarin anticoagulants remain the preferred method of prophylaxis and treatment for thrombosis in a range of clinical situations. Around 1% of the populations of many countries receive such treatment, a proportion boosted by the increasing use of oral anticoagulation as thromboprophylaxis in atrial fibrillation.2 Monitoring of the anticoagulant effect of coumarins, however, is essential, and the prothrombin time is used for this …Keywords
This publication has 8 references indexed in Scilit:
- Reliability of international normalised ratios from two point of care test systems: comparison with conventional methodsBMJ, 2003
- A randomised controlled trial of patient self management of oral anticoagulation treatment compared with primary care managementJournal of Clinical Pathology, 2002
- Reliability of point‐of‐care prothrombin time testing in a community clinic: a randomized crossover comparison with hospital laboratory testingBritish Journal of Haematology, 2002
- Near‐patient testing devices to monitor oral anticoagulant therapyBritish Journal of Haematology, 2001
- A primary care evaluation of three near patient coagulometers.Journal of Clinical Pathology, 1999
- Is the international normalised ratio (INR) reliable? A trial of comparative measurements in hospital laboratory and primary care settingsJournal of Clinical Pathology, 1999
- Accuracy of a Portable International Normalization Ratio Monitor in Outpatients Receiving Long-Term Oral Anticoagulant TherapyThrombosis Research, 1998
- International normalized ratio for prothrombin times in patients taking oral anticoagulants: critical difference and probability of significant change in consecutive measurementsClinical Chemistry, 1995