A comparison of beam characteristics for gated and nongated clinical x-ray beams
- 1 October 1999
- journal article
- research article
- Published by Wiley in Medical Physics
- Vol. 26 (10) , 2086-2091
- https://doi.org/10.1118/1.598723
Abstract
Respiratory gating has only recently been applied to conventional external beam radiotherapy. In order for respiratory gating to be used clinically, an evaluation of the dosimetric effects of small units of delivered dose must be performed. The purpose of this study is to systematically evaluate the effect of various gating sequences on x-ray central axis output, ionization ratios (nominal accelerating potential), beam flatness, and beam symmetry. Measurements were taken for 6 and 18 MV photons on a linear accelerator that generates the gate by using a gridded electron gun to stop the electron flow to the wave-guide. The beam output, energy, flatness, and symmetry did not vary by more than 0.8 percent in most of the gating sequences. The maximum output deviations (0.8 percent), flatness deviations (1.9 percent), and symmetry deviations (0.8 percent) occurred when a low number of monitor units (<5 MU) were delivered in the gating window. Although these deviations are not clinically significant, each linear accelerator should be evaluated carefully before clinical implementation.Keywords
This publication has 22 references indexed in Scilit:
- Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapyMedical Physics, 1998
- Radiotherapy Alone for Medically Inoperable Stage I Non-Small–Cell Lung Cancer: The Duke ExperienceInternational Journal of Radiation Oncology*Biology*Physics, 1998
- Output stability of a linear accelerator during the first three secondsMedical Physics, 1996
- Respiration gated radiotherapy treatment: a technical studyPhysics in Medicine & Biology, 1996
- Ultrasound quantitation of respiratory organ motion in the upper abdomenThe British Journal of Radiology, 1994
- Accurate contiguous sections without breath-holding on chest CT: value of respiratory gating and ultrafast CT.American Journal of Roentgenology, 1994
- The influence of respiration induced motion of the kidneys on the accuracy of radiotherapy treatment planning, a magnetic resonance imaging studyRadiotherapy and Oncology, 1994
- A monitoring, feedback, and triggering system for reproducible breath‐hold MR imagingMagnetic Resonance in Medicine, 1993
- Respiratory kinematics of the upper abdominal organs: A quantitative studyMagnetic Resonance in Medicine, 1992
- Irradiation synchronized with respiration gateInternational Journal of Radiation Oncology*Biology*Physics, 1989