Autologous blood collection in abdominal vascular surgery. Assessment of a low pressure blood salvage system with particular reference to the preservation of cellular elements, triglyceride, complement and bacterial content in the collected blood
- 28 June 2008
- journal article
- Published by Wiley in Clinical and Laboratory Haematology
- Vol. 5 (4) , 361-370
- https://doi.org/10.1111/j.1365-2257.1983.tb00509.x
Abstract
A simple, low pressure, blood scavenging system has been assessed in major abdominal vascular operations. The aim of this study was to assess the quality of blood obtained with this device and its suitability for reinfusion. Three areas of special interest, which have not been reported so far, are the fate of plasma complement, the load of fat aspirated after tissue dissection and the degree of bacterial contamination in the scavenged blood. The development of autologous blood scavenging systems is reviewed. The features which most affect the quality of scavenged blood are identified and their importance discussed by comparison with our experience. Key features of our system were: simplicity of the apparatus, controlled low-pressure aspiration, the use of systemic heparin and the avoidance of mechanical pumps. The blood obtained was of excellent quality with good preservation of cellular elements, platelets and fibrinogen. Plasma total haemolytic complement, C3 and C4 fractions were preserved in normal, though slightly reduced, quantities. Lipid (triglyceride) content was minimally increased after filtration. Bacterial contamination was present in all cases, but at a very low level provided that aspiration was limited to the peritoneal cavity. This low-level of contamination is not thought to be of great significance; its origin and importance are discussed.Keywords
This publication has 11 references indexed in Scilit:
- Does a Relationship Exist between Massive Blood Transfusions and the Adult Respiratory Distress Syndrome? If so, What are the Best Preventive Measures?Vox Sanguinis, 1977
- Intraoperative autotransfusion in the management of massive hemorrhageThe American Journal of Surgery, 1975
- Intraoperative autologous blood transfusion: report of a technique.Thorax, 1975
- Pulmonary Microembolism Associated With Massive TransfusionAnnals of Surgery, 1975
- Intraoperative Autotransfusion in Abdominal Aortic ResectionsArchives of Surgery, 1973
- AUTOTRANSFUSION FROM HEMOTHORAXPublished by Wolters Kluwer Health ,1972
- Early Clinical Experience with a Disposable Unit for the Intraoperative Salvage and Reinfusion of Blood Loss (Intraoperative Autotransfusion)The American Journal of Surgery, 1970
- AUTOTRANSFUSION IN RUPTURED ECTOPIC PREGNANCYThe Lancet, 1970
- On Re-Infusion of Blood in Primary and Other AmputationsBMJ, 1886
- PRACTICAL REMARKS ON AN OVERLOOKED SOURCE OF BLOOD-SUPPLY FOR TRANSFUSION IN POST-PARTUM HÆMORRHAGE, SUGGESTED BY A RECENT FATAL CASE.The Lancet, 1874