Intensified Ultraviolet Analyzing Lamp for Flash Photolysis
- 1 December 1972
- journal article
- research article
- Published by AIP Publishing in Review of Scientific Instruments
- Vol. 43 (12) , 1797-1799
- https://doi.org/10.1063/1.1685568
Abstract
An intensifier is described which increases the light level of an Osram 150 W xenon lamp by a factor of up to 30 at 350 nm, and holds it constant to 0.02% of the brightened level for 10 msec. Control for longer times is possible. This precision of control is possible only when the light level seen by a photomultiplier is used to regulate the intensified level. If two photomultipliers are used, one to observe the transient absorption signals and the other to control the light level, then the light level can be held constant to about 0.1%. This limitation is not due to the intensifier circuit, but to the difficulty of optically matching two beams and two photomultipliers.Keywords
This publication has 7 references indexed in Scilit:
- An Intense Microsecond Flash LampReview of Scientific Instruments, 1972
- Some Comments on Xenon Arc Lamp StabilityReview of Scientific Instruments, 1972
- Some Characteristics of a Pulsed Xenon Lamp for Use as a Light Source in Kinetic SpectrophotometryReview of Scientific Instruments, 1972
- FLASH PHOTOLYSIS STUDIES OF OROTIC ACIDPhotochemistry and Photobiology, 1971
- Properties of the triplet states of thymine and uracil in aqueous solutionJournal of the American Chemical Society, 1971
- Optical detection of the triplet state of uracilBiochemical and Biophysical Research Communications, 1969
- Pulse Radiolysis Apparatus for Monitoring at 2000 ÅReview of Scientific Instruments, 1969