Correlation of Spectral Phonoangiography and Carotid Angiography with Gross Pathology in Carotid Stenosis
- 20 August 1981
- journal article
- research article
- Published by Massachusetts Medical Society in New England Journal of Medicine
- Vol. 305 (8) , 417-419
- https://doi.org/10.1056/nejm198108203050801
Abstract
Spectral phonoangiography, a noninvasive method for measurement of the residual-lumen diameter of carotid stenosis by bruit analysis, was compared with x-ray angiography and direct measurement of the pathological specimen at carotid endarterectomy in 39 bifurcations from 36 patients. In six studies, the bruit was too faint to analyze. In 31 of the other 33 studies, the phonoangiogram predicted the residual-lumen diameter to within 0.5 mm of the measured value. Of the 39 contrast angiograms, 35 showed residual lumens within 0.5 mm of the value measured in the specimen, two showed lumens between 0.5 and 1 mm, and the sizes of two could not be estimated because of vessel overlap in all planes. Spectral phonoangiography and contrast angiography are both accurate methods for evaluation of carotid stenosis. Since phonoangiography is noninvasive, it may be of particular value in determining the natural history in patients with carotid bruits. (N Engl J Med. 1981; 305:417–9.)This publication has 6 references indexed in Scilit:
- Effects of surrounding tissue on the sound spectrum of arterial bruits in vivo.Stroke, 1980
- Changes in internal carotid artery flow velocities with cerebral vasodilation and constriction.Stroke, 1979
- The bruit of carotid stenosis versus radiated basal heart murmurs. Differentiation by phonoangiography.Circulation, 1978
- Evaluation of Carotid Stenosis by PhonoangiographyNew England Journal of Medicine, 1975
- ATHEROSCLEROSIS OF THE CAROTID AND VERTEBRAL ARTERIES—EXTRACRANIAL AND INTRACRANIALJournal of Neuropathology and Experimental Neurology, 1965
- OCCLUSION OF THE INTERNAL CAROTID ARTERYArchives of Neurology & Psychiatry, 1951